AC-DS-03 Automated Deactivation System Manual Ver. A03

Principle of Operation

The AC-DS-03 is a semi automatic deactivation station used in the disposal of high level
disinfectants. The operator manually adds the proper quantity of neutralizer and the unit
automatically pumps the neutralized disinfectant to drain when full. The unit can be activated to
pump the tank at any time with a manual purge feature. The AC-DS-03 can be used in
conjunction with automatic processing devices for endoscopic ultrasound probes.

Safety Precautions

Disconnect power before replacing or servicing the unit.
Use PPE whenever operating the AC-DS-03 — wear safety goggles, lab coat, proper gloves and
face/eye shield.

Environmental Conditions

For indoor use only.
Temperature range: 5C to 40C.
Main supply voltage fluctuations not to exceed +/- 10% of the nominal voltage.
Insulation over voltage category II.
Pollution degree 2.

Supply Ratings

Voltage:
Current draw:
Fuse:
AC 110-120V 0.5A 50/60 Hz
50 watts
1 each Type FSF 5 x 20 mm 2 Amp 250VAC externally panel mounted.
1