Output Cable - Sony DG110BM Instruction Manual

Digital gauge
Table of Contents

Advertisement

Available languages

Available languages

6-4. Output Cable

Use a shielded cable as shown in the figure below for output signal. Connect the shielded
wires of the cable to the case of each connector at both ends of cable.
Make the cable length as short as possible to prevent noise.
. When using a power supply with remote sensing function
Use a twisted pair with a thickness of at least 26 AWG for the output cable. The output
signals use voltage-differential line driver output.
Connect the shielded wires to FG.
Set the supply voltage so that it satisfies the specified value at the interpolator input.
Recommended cables : 20276-VSV-10P x 26AWG-7/0.16 (Hirakawa Hewtech)
Reference
Remote sensing function
By connecting the sensing pins (+S, _S), the voltage at point a (see the following page)
is automatically maintained at the voltage value set on the power supply unit side even if
the output cable is lengthened (max. 50 m). So, there is no need to take into account the
voltage drop at the cable.
<Example power supply with remote sensing function> Densei Lambda NNS Series
. When using a power supply without remote sensing function
Use a twisted pair with a thickness of at least 20 AWG for the output cable.
Set the supply voltage so that it satisfies the specified value at the interpolator input.
DG110BM/DG110BE
Outer sheathe
Braided shielding
11
(E)

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dg110be

Table of Contents