7
Restore Factory Settings
Press the Enter button, press the Select button, then press Adjust button to input
the password 6100. Press the Enter button to display FFFF, which indicates that the
system has been restored to the factory setting. Press the Enter button to return to
the main interface.
NoTE: The temperature unit and voice volume is not included in the factory reset
parameters.

Installation Guidelines

Install the device indoors, as wind and sunshine can interfere with the thermo
detection.
Thermo Detection Notes
1.
The normal use environment temperature is 32° F ~ 86° F. If you are using when
temperature is approximately -14° F, the ambient temperature compensation
value needs to be increased by 5°, until the error between the measured
value and the forehead gun is within ± 0.3°. If multiple body temperature
measurements are abnormally low, you can also use this method to calibrate the
detector.
2.
The best measurement distance is .393 ~ 3 inches.
3.
The measurement time is 0.5 seconds. To avoid repetitive measurement, please
move forehead or hand away immediately after detection.
4.
The measurement accuracy is ±0.3°.