Cell Mode; Data Log Mode - Martek MARK 22 Operation Manual

In situ calibrator
Table of Contents

Advertisement

in regards to operation of the Mark 22.
To select a SETUP option, the user must
depress the SELECT key. To move the flash-
ing prompt up or down to another option,
you would press either the UP or DOWN
arrow keys. To exit from the Setup Option
Menu the user must depress the EXIT key.
which will return the Mark 22 to the normal
display monitoring mode. The options pre-
sented in the SETUP menu are discussed in
Section 4.2 and in the chapter on CALIBRA-
TION. For now, simply depress the EXIT key.
In the Setup Mode of operation the top row
of keys function as indicated by the yellow
text located on the top line of text on each of
the function keys.

4.1.3 Cell Mode

The CELL Key allows you to change the
cell constant currently used by the Mark 22
to any of three different, preset cell constants
simply by pressing the appropriate function
key whose cell constant value is labeled in
blue. Press the CELL key. The Mark 22 will
respond with a display similar to the one pic-
tured below:
CELL #1
CELL #3
To change this cell constant, simply press
one of the function keys. Pressing a function
key while in the Cell Mode will change the
K factor used to calculate the raw and tem-
perature-compensated
surements. While the nominal cell constant
values listed on the function keys are 0.01,
0.1, 1, and 10, you can change any of these
cell constants to whatever value is desired.
Refer to Section 4.2 on how to change indi-
vidual cell constants. To return to the moni-
toring mode, press the CELL key again. In
the Cell Mode of operation the top row of
Martek Instruments, Inc.
CELL #2
CELL #4
conductivity
mea-
OPERATION - 4
keys functioon as indicated by the blue text
located on the second line of text on each of
the function keys. The K 0.01 Key selects the
cell constant for Cell #1, the K 0.1 Key se-
lects the cell constant for Cell #2, the K 1 key
selects the cell constant for Cell #3 and the K
10 Key selects the Cell constant for Cell #4.
The Cell Screen appears as follows:
CURRENT CELL #1
IS K .01000

4.1.4 Data Log Mode

In the Log Mode of operation the top row
of keys function as indicated by the red text
located on the third line of text on each of
the function keys. The FREE Key will dis-
play the amount of storage memory remain-
ing in terms of hours. The amount of data
stored is directly related to the scan rate used
during the logging operation. The PAUSE
Key will suspend the logging function, and
the RESUME Key will restart it. The CLEAR
Key will clear all of the logged data from the
unit's internal memory. The LOG key is one
of two operational mode keys on the Mark
22 membrane panel outlined in RED. When
the log Key is depressed, the function ( la-
beled in red ) on each of the function keys
becomes applicable. After the LOG key is de-
pressed, the Mark 22 will display the follow-
ing screen:
CURRENT LOG FILE
IS FILE # 01
15

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Martek MARK 22

Table of Contents