Orbital Shaker ..............................15 Shaker Installation ..............................15 Shaker Control ............................... 15 AlgaeTron AG 230 Control ..........................18 Explanation of Symbols and Color Differentiation Used in the Graphical Presentation ........18 Control Menu Tree ..............................20 Gas Input ................................29 Technical Data ..............................
• Do not remove the plastic drain tray fitted above the compressor. • Let the AlgaeTron AG 230 stand up after the transport and WAIT AT LEAST 12 HOURS before connecting it with the electricity. • Use only the cables supplied by the manufacturer.
Page 4
Device does not have to be turned off. WARNING: The AlgaeTron AG 230 is considered Class 1M* LED Product. LED radiation may be harmful to eye, avoid direct and strongly reflected exposure. It is recommended to use protective glasses. *Class 1M: Laser and LED equipment that is safe under reasonable conditions of operation for use with the naked eye.
AlgaeTron AG 230 features three electrical outlets (EU type) for attachment of up to three dedicated orbital shakers SHK-2013. Use of different types of shakers is not recommended!
DEVICE DESCRIPTION AND INSTALLATION Standard version of the AlgaeTron AG 230 package consists of a main body of AG 230 containing control unit to regulate temperature and illumination. All controlled parameters can be set via a small display screen on the top of AlgaeTron front size.
Page 7
Temperature can be set in the range of +15 to +45 (with maximum illumination). The controlled range can be °C customized and extended: from +10 to +55 °C. To allow homogenous temperature distribution inside of the AlgaeTron, inner airflow fans are in-built into the chamber. In total 0.46 m of illuminated cultivation area is available on three slide-out shelves.
COMPONENTS OF AG 230 Please, find bellow a list of standard AlgaeTron AG 230 components delivered to the customer. Note: Check the contents of the package and compare it with enclosed standard package list (see below). LIST OF STANDARD AG 230 COMPONENTS 1.
AlgaeTron AG 230 top. • If you place the rear of the AlgaeTron AG 230 against the wall, be sure there is at least 15 cm to ensure sufficient distance from the wall for heat removal from the condenser.
Used to move up in the menu or to increase value of parameter. [▼] Used to move down in the menu or to reduce value of parameter. See Chapter AlgaeTron AG 230 Control on page 16 for more information about the AlgaeTron Control Panel operation.
DESCRIPTION OF THE ALGAETRON REAR PANEL AND CONNECTION TO 230 OR 110 V POWER SUPPLY The rear panel houses connectors for all connecting cables (Fig.4). Fig.4 The AlgaeTron rear panel. [6] ON/OFF power switch. [7] Firmware communication connector. [8] Gas ports. [9] Main power connectors.
Fig.5 [11] Main cable 230 V. [12] Main cable 110 V. DESCRIPTION OF THE ALGAETRON INTERIOR SPACE The components installed in the upper part of the interior back-wall are shown and described in the Fig.6. Fig.6 The back-wall of the AlgaeTron interior. [13] Upper LED light shelf connector. [14] Airflow Fans. [15] The upper LED panel. [16] Air inlet.
[19] Inside electrical sockets are intended for connection of supplementary accessories (shakers SHK-2013, etc.) inside the AlgaeTron AG 230. The outlet has a fuse, which is situated on the rear panel of the device (see feature [10] in Fig.4). To connect supplementary accessories to the outlet, use only the cable supplied by the manufacturer.
Page 14
Fig.7 The AlgaeTron shelving and inner space compartmentalization. [21], [22] and [23] Cultivation shelves. [24] Two lower LED panels.
AlgaeTron damage and loss of warranty. SHAKER INSTALLATION Place the shaker into the AlgaeTron AG 230 and plug it into the inner electrical socket [19]. Do not overload the shaker! Maximum is 5 kg. Loading capacity:...
Page 16
Fig.9 The position of the control buttons on the orbital shaker. The control panel includes following controls: 2-pole master switch (toggle-type, lighted green on the left side of the unit) Setting knob, shaking frequency Set Time key : time up Set Time key : time down Select key: display selector: shaking frequency (Act Speed) / time (Act Time)
Page 17
• Hitting the Select key during timer-controlled operation, you may change display from time to shaking frequency and vice versa. • Speed LED turns ON. Timer LED flashes to indicate timer operation mode. • Timer LED also flashes, when timer-controlled operation was discontinued with the Start / Stop key. REPETITION •...
The following pages show a graphical representation of the operation scheme for the AlgaeTron. This scheme is structured in five levels. Individual levels are marked with a different color for an easier orientation in the scheme. Note: The AlgaeTron AG 230 front panel does not reflect this color differentiation Main menu...
Page 19
See pages 20 - 28 of this Manual for the graphical representation of individual menus and for explanations of their functions. [M] key: Used to move back in the menu tree or to exit the menu. [S] key: Used to move forward in the menu tree or to save your selection. [↑] key: Used to move up in the menu or to add value.
GAS INPUT AG 230 allows to connect both pressurized and non-pressurized air source. At the rear side of the device, there are two gas input connectors (see Fig.4, feature [8]): • Gas Input 1 is used when the air source is not pressurized, for example outside room air. Selection of the currently used gas input is made in the device menu (Settings >...
Page 30
range 0 – 2 500 mL/min. The user can define the required gas mixture either by setting the flows of the individual gasses (e.g., 980 mL/min of N and 20 mL/min of CO ) or by setting the required relative composition of gas mixture (2% CO ) and the total mixture gas flow (1 000 mL/min).
TECHNICAL DATA ALGAETRON AG 230 Temperature range (ºC) +15…+45 Optionally +10…+55°C (the increased temperature exclude the use of increased customized light intensity) LED Light Illumination (cm), upper panel 25 x 35 LED Light Shelves Illumination (cm) 33.5 x 46.5 LED illumination...
PSI (unless otherwise stated). • If at any time within this warranty period the instrument does not function as warranted, return it and PSI will repair or replace it at no charge. The customer is responsible for shipping and insurance charges (for the full product value) to PSI.