Page 1
O & M Manual GasSens Bromine Gas Monitor Home Office European Office Analytical Technology, Inc. ATI (UK) Limited 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park Collegeville, PA 19426 Delph New Road, Delph Phone: 800-959-0299 Saddleworth OL3 5DE...
GasSens Gas Detection System Section 1 - Overview TABLE OF CONTENTS TABLE OF CONTENTS ............................ 1 INTRODUCTION ............................1-3 A11 S ..........................1-4 ENSOR RANSMITTER A14 R ..........................1-4 ECEIVER ODULE A17 P ......................... 1-4 OWER UPPLY ODULE NEMA 4X E ..........................
Page 3
GasSens Gas Detection System Section 1 - Overview BATTERY BACKUP PARTS LIST ....................... 4-3 STROBE PARTS LIST ............................ 4-3 STROBE LIGHT ............................. 4-3 O & M Manual A14-RK, 8/06 1 - 2...
0-1000 PPB 0-100 PPM * Auto-Test is available in these gases. The GasSens system consists of modular components that can be used in a variety of configurations to fit specific application requirements. A description of the major system components follows: O &...
GasSens Gas Detection System Section 1 - Overview A11 Sensor/Transmitter Consists of an electrochemical gas sensor closely coupled to a NEMA 4X transmitter. This component measures gas concentration and converts the measurement to a digital signal for transmission to a receiver module. It must be mounted in the area where gas leakage or buildup is expected, and can either be wall mounted or supported by electrical conduit carrying interconnecting wiring.
If AC power is lost, the battery insures that detection continues uninterrupted. GasSens components allow simple assembly of either single or multipoint detection systems. Each point of detection requires one sensor/transmitter and one receiver.
GasSens Gas Detection System Section 1 - Overview MECHANICAL INSTALLATION System enclosures, battery back-up units, and sensor transmitters are all designed for surface mounting using screws or bolts inserted through the recessed mounting holes at each corner of the enclosures. Included with each system enclosure is a template with mounting hole dimensions for two and three module system enclosures, battery back-up enclosure, and sensor/transmitter enclosure.
Page 8
GasSens Gas Detection System Section 1 - Overview For installations where conduit will not be used, cable gland seal fittings must be installed in the conduit hubs. Plastic cable glands are available from most electrical supply stores. Gland seals with 1/2" NPT threads will screw directly into the conduit hub. Gland seals are also available from ATI for installation directly into the enclosures in place of the conduit hubs.
Page 9
GasSens Gas Detection System Section 1 - Overview FRONT VIEW SIDE VIEW THRU HOLES FOR SURFACE MOUNTING (4) #6 SCREWS BOTTOM VIEW Figure 1-2: Two Module System Enclosure Dimensions (80-0006R) O & M Manual A14-RK, 8/06 1 - 8...
Page 10
GasSens Gas Detection System Section 1 - Overview FRONT VIEW SIDE VIEW THRU HOLES FOR SURFACE MOUNTING (4) #6 SCREWS BOTTOM VIEW Figure 1-3: Three Module System Enclosure Dimensions (80-0007R) O & M Manual A14-RK, 8/06 1 - 9...
Page 11
GasSens Gas Detection System Section 1 - Overview Figure 1-4: Six Module System Enclosure Dimensions (80-0008R) O & M Manual A14-RK, 8/06 1 - 10...
Receiver module. Figure 3-1 in section 3 shows a detailed terminal wiring drawing for the power supply module. A GasSens gas detection system will always contain an A14 Receiver and an A11 Sensor/ transmitter. Normally, an A17 Power Supply module will also be included, and an A19 Battery Back-up Unit will often be used to provide standby power for the system.
Page 13
GasSens Gas Detection System Section 1 - Overview Two point detectors include one power supply and two receivers. These systems are shipped factory wired as shown in Figure 1-6 below. Analytical Technology, Inc. Figure 1-6: Dual Point Receiver Internal Wiring, (ATI-033) O &...
Page 14
GasSens Gas Detection System Section 1 - Overview Analytical Technology, Inc. Analytical Technology, Inc. Figure 1-7: Four Point Receiver Internal Wiring (ATI-075) O & M Manual A14-RK, 8/06 1 - 13...
Page 15
GasSens Gas Detection System Section 1 - Overview Multipoint systems supplied in “Large” type enclosures use a 65 watt power supply located in the top of the enclosure. This power supply is designed to power up to 12 of the A14 receiver modules.
Page 16
GasSens Gas Detection System Section 1 - Overview Sensor/transmitters contain two 2-position terminal blocks, one for receiver connection and one for connection of the Auto-Test generator. Connection of the two wires from the receiver are not polarity sensitive. These wires can be hooked up without regard to terminal position.
Page 17
GasSens Gas Detection System Section 1 - Overview Analytical Technology, Inc. Figure 1-10: Typical System Wiring, Explosion-proof Transmitter Version (ATI-076) The remainder of this manual is divided into sections that describe each component of the system. Mechanical and electrical installation are the only requirements to provide a workable system.
GasSens Gas Detection System Section 2 - A14 Receiver Module RECEIVER MODULE Introduction Series A14 Receivers provide the user interface to the gas detection system. Each receiver is connected to a series A11 sensor/transmitter with a 2 conductor cable, and displays information on the gas concentration in the sensor area.
Page 19
GasSens Gas Detection System Section 2 - A14 Receiver Module TROUBLE 1: +12 Module Power positive (12 VDC) 1: A1 NO Alarm 1 normally open contact 2: C Module Power Common 2: A1 C Alarm 1 common 3: Earth Ground (REQUIRED)
GasSens Gas Detection System Section 2 - A14 Receiver Module Factory Configuration All receivers are identical except for the gas symbol attached to the green power LED. Operating range, alarm relay configuration, display intensity, and Auto-Test activation are selectable using 4 banks of DIP switches located on the configuration circuit board inside the module.
Page 21
GasSens Gas Detection System Section 2 - A14 Receiver Module Figure 2-4: Configuration Switch Functions (0110PM) O & M Manual A14-RK, 8/06 2 - 4...
GasSens Gas Detection System Section 2 - A14 Receiver Module Setpoint Selection The operating range and the Warning and Alarm setpoints can be changed, if desired, by using the configuration switches identified in Figure 2-4. The Warning setpoint is set using switches A2-A8 and the Alarm setpoint is set using switches B2-B8.
GasSens Gas Detection System Section 2 - A14 Receiver Module Range Selection Display range selection is made using binary numbers to represent the full scale operating range for the receiver. Table 2-2 provides the proper switch settings for the ranges available on each receiver.
GasSens Gas Detection System Section 2 - A14 Receiver Module Relays may be configured for either Fail-safe or Normal operation. The designation of normal or fail-safe refers to the operation of the relay coil during normal operation. A relay configured for Normal operation will have the relay coil energized when an alarm occurs and de-energized during normal operation.
GasSens Gas Detection System Section 2 - A14 Receiver Module OPERATION Startup Receiver front panels contain four LED bar indicators and a 4 digit LED display. The digital display indicates gas concentration in either PPM, PPB, or percent. The green LED bar is marked with the gas symbol and units of measurement (PPM, PPB, or %) specified on the customer order.
GasSens Gas Detection System Section 2 - A14 Receiver Module Trouble Alarm and Relay Receivers contain a TROUBLE indicator and an associated SPDT relay. The trouble alarm will be activated if the transmitter cable is broken or shorted, or if electronic component failure in either the transmitter or receiver causes a loss of digital input.
GasSens Gas Detection System Section 2 - A14 Receiver Module When the Auto-Test is activated, the generator will turn on for a maximum of 5 minutes. If no sensor response is detected during this time, the generator will turn off and the receiver will wait for 15 minutes before initiating another Auto-Test.
GasSens Gas Detection System Section 2 - A14 Receiver Module Manual Auto-Test Gas detection systems purchased with the optional Auto-Test generator may be tested by manual activation of the generator from the receiver module. This function will operate only if the Auto-Test configuration switch D1 is 'on'.
GasSens Gas Detection System Section 2 - A14 Receiver Module Auto-Test Remote Verification A14 Receiver modules with a software revision G or higher use the analog output to indicate that a successful Auto-Test has occurred. This allows a recording device to show evidence of the successful test.
GasSens Gas Detection System Section 2 - A14 Receiver Module TROUBLESHOOTING Receiver modules will normally provide trouble free operation over many years of service. However, should problems arise, a few simple tests can be done to determine if the receiver is functioning properly.
GasSens Gas Detection System Section 2 - A14 Receiver Module RECEIVER MODULE PARTS LIST TOP HALF OF ENCLOSURE (80-0002) PLUG-ON TERMINAL BLOCK (38-0006) BOTTOM HALF OF ENCLOSURE (80-0002) PC BOARD ASSEMLBY FRONT PANEL (34-0009) PLUG-ON TERMINAL BLOCK (38-0009) Part Number...
Section 3 – A17 Power Supply Module A17 UNIVERSAL POWER SUPPLY The power supply module used in the GasSens system (part #00-0055) is a self-adjusting supply that will accept the AC power provided in virtually every country in the world. Any AC (50-60 Hz.) or DC voltage from 85 volts to 270 volts can be connected to the power input terminals on TB4.
Page 33
Universal Power Supply Section 3 – A17 Power Supply Module Figure 3-1 provides detailed information on the terminals provided on the power supply. All terminal blocks are plug-in type, and can be easily unplugged should removal of the module be necessary. Analytical Technology, Inc.
Universal Power Supply Section 3 – A17 Power Supply Module POWER SUPPLY PARTS LIST Part Number Description 00-0055 Complete power supply module 81-0064 Module enclosure (top and bottom) 34-0011 Front overlay 38-0065 AC power terminal block plug, 3 position 38-0066 Power failure relay &...
Section 4 - Optional Equipment OPTIONAL EQUIPMENT Battery back-up units for GasSens detection systems (part #00-0057) are separate components housed in a NEMA 4X wall mount enclosure. Back-up units include a 4 ampere-hour sealed lead acid battery with a control circuit attached to the battery terminals. The battery is held in place on an aluminum bracket fixed to the plastic enclosure.
GasSens Gas Detection System Section 4 - Optional Equipment Electrical connection is made between the battery back-up unit and the power supply module. The circuit board mounted on the battery contains a 2 position plug-in terminal block marked plus and minus.
Page 37
GasSens Gas Detection System Section 4 - Optional Equipment BATTERY BACKUP PARTS LIST Part Number Description 00-0057 Complete battery back-up unit 01-0011 Battery back-up circuit board 80-0009 NEMA 4X enclosure (top and bottom) 48-0010 Battery bracket 29-0002 Battery, 12 V, 4 A-H 92-0022 Self-tapping screws, (Pkg.
Page 38
Section 4 - Optional Equipment STROBE LIGHT The accessory alarm indicating strobe light (part number 35-0002) available for use with the GasSens alarm system is a weatherproof high-intensity strobe operating from a 12 VDC power source. The strobe can be conveniently mounted on the top or either side of the alarm enclosure using one of the 1/2" NPT hubs supplied with the unit.
Page 39
GasSens Gas Detection System Section 5 - A11-10 Br 2 Sensor/Transmitter INTRODUCTION Series A11 sensor/transmitters combine electrochemical gas sensors and an electronic amplifier that transmits gas concentration using a current pulse position technique. When ordered as an option, sensor/transmitters are also supplied with the gas generator that provides the Auto-Test automatic sensor testing system.
Page 40
GasSens Gas Detection System Section 5 - A11-10 Br 2 Sensor/Transmitter Notes: 1) Enclosure Ratings: Nema-4X/IP 66. 2) Enclosure Material: Polystyrene, standard gray 3) Knockouts: Pg 11 (.75” dia.) Pg 16 (.90” dia.) 4) **These are the dimensions between knockouts and are typical for (3) sides with (2) knockouts each.
Page 41
GasSens Gas Detection System Section 5 - A11-10 Br 2 Sensor/Transmitter Figure 5-2: Explosion-Proof Sensor/Transmitter Dimensions (ATI-046, ATI-0170) O & M Manual Revision E, 10/15 5 - 3...
Page 42
GasSens Gas Detection System Section 5 - A11-10 Br 2 Sensor/Transmitter SENSOR LOCATION Bromine gas is heavier than air and will tend to accumulate near the floor in a closed room with little air movement. If there is good air circulation in an enclosed area, some bromine will undoubtedly be carried throughout the area.
Page 43
GasSens Gas Detection System Section 5 - A11-10 Br 2 Sensor/Transmitter CALIBRATION Sensor/transmitters are factory calibrated for the operating range specified on the order, with the calibrated range shown on the calibration tag attached inside the lid of the transmitter enclosure. For critical applications, sensor/transmitters should be calibrated every 3-6 months.
Page 44
GasSens Gas Detection System Section 5 - A11-10 Br 2 Sensor/Transmitter FIGURE 5-2: SENSOR/TRANSMITTER CONTROLS AND TEST POINTS (ATI-096) TRANSMITTER ZERO The transmitter zero is adjusted with the sensor exposed to air that contains no bromine. Generally, the easiest method of zeroing the transmitter is to make the adjustment when you know that the area is free of bromine.
Page 45
GasSens Gas Detection System Section 5 - A11-10 Br 2 Sensor/Transmitter TRANSMITTER SPAN To span the sensor/transmitter, select the calibration generator output range of 1 PPM and allow the generator to warm up for approximately 30 minutes. Connect the output tube from the standards generator to the sensor calibration adapter.
Page 46
GasSens Gas Detection System Section 5 - A11-10 Br 2 Sensor/Transmitter SENSOR REPLACEMENT Electrochemical sensors used in the A11 are warranted for 12 months and generally last 18-24 months or more. When sensor replacement is required, it can be done easily and quickly. Open the transmitter and unplug the sensor cable from the transmitter circuit board.
Page 47
GasSens Gas Detection System Section 5 - A11-10 Br 2 Sensor/Transmitter Br 2 SENSOR/TRANSMITTER PARTS LIST Part Number Description 00-0119 Complete bromine sensor/transmitter assembly (including sensor) 00-0002 Bromine gas sensor 00-0097 Chlorine gas generator with gland seal (used to Auto-Test Br 2 systems)
Page 48
PRODUCT WARRANTY Analytical Technology, Inc. (Manufacturer) warrants to the Customer that if any part(s) of the Manufacturer's equipment proves to be defective in materials or workmanship within the earlier of 18 months of the date of shipment or 12 months of the date of start- up, such defective parts will be repaired or replaced free of charge.
Need help?
Do you have a question about the GasSens and is the answer not in the manual?
Questions and answers