Alkalinity
Hide thumbs Also See for HI775:

Advertisement

Quick Links

Tips for an Accurate Measurement
• It is important that the sample does not contain any debris.
• Whenever the cuvette is placed into the measurement cell, it
must be dry outside, and completely free of fingerprints, oil and
dirt. Wipe it thoroughly with
HI731318
or a lint-free cloth prior
to insertion.
• Shaking the cuvette can generate bubbles, causing higher
readings. To obtain accurate measurements, remove such bubbles
by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand for too long after reagent
is added, as accuracy will be affected.
• After the reading it is important to immediately discard the
sample, otherwise the glass might become permanently stained.
Battery Management
To save the battery, the instrument shuts down after 10 minutes of non-use and 2 minutes
after reading.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display "bAd" then "bAt" for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument's battery:
• Turn the instrument off by holding the button until the meter shuts off.
• Turn the instrument upside down and remove the battery cover with a screwdriver.
• Remove the battery from its location and replace it with a new one, inserting the negative
end first.
• Insert the battery cover and replace the screw with a screwdriver.
Recommendations for Users
Before using Hanna Instruments products, make sure that they are entirely suitable for your
specific application and for the environment in which they are used. Any variation introduced
by the user to the supplied equipment may degrade the instrument's performance. For yours
and the instrument safety do not use or store the instrument in hazardous environments.
Accessories
Reagent Sets
HI775-26
Reagents for 25 Alkalinity tests, 1 syringe and 1 tip
Other Accessories
HI775-11
Alkalinity Certified Standard Kit
HI731318
Cloth for wiping cuvettes (4 pcs.)
HI731321
Glass cuvettes and Seal Cap for Checker
HI731225
Cuvette Black Cap for Checker
HC Colorimeters (4 pcs.)
®
HI731353
Cuvette seal cap for Checker
HC Colorimeters (4 pcs.)
®
HI740028P
1.5V AAA batteries (12 pcs.)
HI740142P
1 mL graduated syringe with tip (10 pcs.)
HI93703-50
Cuvette cleaning solution (230 mL)
HI93755-53
Chlorine Remover
Warranty
HI775
is warranted for a period of one year after date of purchase against defects in
workmanship and materials when used for their intended purpose and maintained according
to instructions. This warranty is limited to repair or replacement free of charge. Damage due
to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service
is required, contact your local Hanna Instruments Office. If under warranty, report the model
number, date of purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Return Goods Authorization number from the
Technical Service department and then send it with shipping costs prepaid. When shipping any
instrument, make sure it is properly packed for complete protection.
INSTRUCTION MANUAL
®
HC Colorimeters (4 pcs.)
IST775 01/19
HI775
Alkalinity

Advertisement

loading
Need help?

Need help?

Do you have a question about the HI775 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hanna Instruments HI775

  • Page 1 Recommendations for Users INSTRUCTION MANUAL Before using Hanna Instruments products, make sure that they are entirely suitable for your • It is important that the sample does not contain any debris. specific application and for the environment in which they are used. Any variation introduced •...
  • Page 2 • Fill the cuvette with 10 mL of unreacted sample Cuvette holder For technical support, contact your local Hanna Instruments Office or e-mail us at and replace the cap. Place the cuvette into the tech@hannainst.com. LCD Liquid Crystal Display meter and close the meter’s cap.