10 . Troubleshooting - Nakanishi AL -C1204 Operation Manual

Air line kit
Table of Contents

Advertisement

9 - 2 Connection of the Air Line Kit
(1) Connect the
6.0mm Air Hose to the Primary Joint.
(2) Connect the
6mm air hose to the Secondary Joint to the Air Input Quick Disconnect on the front of the
CONTROLLER.
(3) Insert the
6mm -
CONTROLLER. Insert the
Disconnect on the back of the motor (
Accessories). (Fig. 6).
Connection Hose
Primary Joint
Connector for Air Hose

10 . TROUBLESHOOTING

If a problem or concern occur, please check the following items prior to consulting your dealer.
Trouble
Broken connection hose and
supply air hose.
No air ow.
Air leakage.
Low air pressure.
Water, dirt and debris are collected
in the Air Filter.
11 . DISPOSAL OF THE AIR LINE KIT
When disposal of an Air Line Kit is necessary, follow the instructions from your local government agency for proper
disposal of electrical components .
4mm Conversion Adaptor into the Air Output Quick Disconnect on the front of the
4mm air hose from the
6mm -
Replace the connection hose and supply air hose.
Check the compressor power supply and the air compressor output.
Check if connection hose and supply air hose is broken, bent or
disconnected.
Check the Regulator and set to the correct air pressure.
Check all connection hose and supply air hose connections.
ON / OFF valve is OFF position. The ON / OFF valve set to ON position.
Check all threaded joints and re-tighten if necessary.
Check the Compressor, Air Circuit, and Regulator.
Drain water, dirt and debris from the Air Filter bowl.
6mm -
4mm Conversion Adaptor to the 4mm Quick
4mm Conversion adaptor : CONTROLLER's Standard
Air Output Joint
Air Input Joint
Secondary Joint
6 One - Touch Joint
Motor (EM-3030J / EM-3060J)
Fig. 6
Inspection / Corrective Active
E3000 CONTROLLER
15