Maintenance
Before Each Use
Technical Inspections
Disposal at the End of Its Service Life
A-4
Cleaning and Maintenance: Maintenance
Before each use, visually inspect the equipment, the cables and the
electrodes for signs of mechanical damage.
If you detect damage or impaired functions which may result in a hazard
to the patient or the operator, the items must be repaired or replaced
before use.
The system does not require other regular maintenance.
For safety, the devices require regular maintenance. To ensure
functional and operational safety of this system, Technical Inspections
should be carried out on an annual basis.
These checks should be referred to individuals with adequate training
and experience.
The inspections can be carried out by GE Healthcare Service within the
framework of a service agreement. The Technical Inspections include:
Visual inspection of the device and of the accessories for mechanical
damage that may impair their function.
Verification of the safety-related labeling for legibility.
A function test.
Measurement of the leakage current and of the resistance of the
protective conductor according to local regulations.
The system does not require other regular maintenance.
This symbol indicates that the waste of electrical and
electronic equipment must not be diposed as unsorted
municipal waste and must be collected separately. Please
contact an authorized representative of the manufacturer
for information concerning the decommissioning of your
equipment.