Page 2
Safet y Platform Walker WARRANTY Wy’East Medical Corporation warrants to the original user that the Whitney Safety Platform Walker manufactured by the company is free from defects in material and workmanship under normal use and service. The term of warranty from the date of purchase/shipment is 2 years on the mechanical lift components, 1 year on the electronic components, and 90 days for the battery.
If anything is not completely understood, please contact your supplier for more details. Failure to comply with warnings in this manual may result in injury. Keep this manual with the Whitney Safety Platform Walker and refer to it as required. Contents of this manual are subjected to change without prior notice to users.
Page 5
Not only does it act as a walker, it also promotes mobility and safety with its unique power rise function and safety harness. The Whitney can be used by anyone who regularly uses a walker. The user can remain active while the power rise function is in use.
1. The Whitney Walker is to be used ONLY on level surfaces. 2. ALWAYS lock the wheels while lifting and lowering the patient. 3. Prior to using the Whitney Walker, the patient must be assessed by a qualified professional to determine the suitability of using the Whitney Walker with that patient.
4. Warning: Check the sling for any fraying, loose stitching, attachments or any deteriorations. 5. After use, check the sling according to steps 3 and 4 and store in a dry area. 52521 R01 Whitney Safety Platform Walker 800.255.3126 | wyeastmedical.com Operator’s Manual...
Depress the foot brake to set it and lock the caster. To unlock, press down on the protruding part of the brake. Locking affects rolling and swiveling. 52521 R01 Whitney Safety Platform Walker 800.255.3126 | wyeastmedical.com Operator’s Manual Page 8 of 20...
Lower 2. To Lower frame use and to go faster legs frame Fast Fast Down 3. To Open Legs use 4. To Close Legs use 52521 R01 Whitney Safety Platform Walker 800.255.3126 | wyeastmedical.com Operator’s Manual Page 9 of 20...
LED flashes green. • Make sure the battery is fully charged before first use. It takes around 24 hours to fully charge the battery. 52521 R01 Whitney Safety Platform Walker 800.255.3126 | wyeastmedical.com Operator’s Manual Page 10 of 20...
The system is working MEDLINE CARELINE user manual • No LED – On battery, not activated by hand or foot control or CA30/CA40 not connected to mains 52521 R01 Whitney Safety Platform Walker 800.255.3126 | wyeastmedical.com Operator’s Manual Page 11 of 20...
WHITNEY Safet y Platform Walker OPERATING THE WHITNEY SAFETY PLATFORM WALKER: 1. Fit the harness as low as possible across the hips and bottom with the broad upper part around the hips and the narrow lower part around the bottom.
Unplug mains cable. wall outlet. Battery, actuators or control Tighten or replace fasteners box are loose Battery will not charge Battery is damaged. Replace battery. 52521 R01 Whitney Safety Platform Walker 800.255.3126 | wyeastmedical.com Operator’s Manual Page 14 of 20...
Page 15
Standing belt slides off belt Walker lockplate broken Replace lockplate holder Standing belt holder is bent Product has been overloaded Replace frame before further use 52521 R01 Whitney Safety Platform Walker 800.255.3126 | wyeastmedical.com Operator’s Manual Page 15 of 20...
4. Use of chemical cleaning for disinfecting may cause material color to fade and sling to wear more rapidly than with typical laundering. Table 1.0 Approved Disinfecting Solutions for Whitney Belt (S 52530, M 52501, L 52502, XL 52529) CLEANER/...
50°C, such as inside a hot automobile, in direct sunlight, or in front of a stove or a source of intense heat. Doing so can shorten battery life, lower its performance level, cause the battery to leak fluid, be damaged or deformed. 52521 R01 Whitney Safety Platform Walker 800.255.3126 | wyeastmedical.com Operator’s Manual...
Do not drop an actuator or otherwise damage the housing during disassembly or transportation. We do not recommend using an actuator which has been damaged. 52521 R01 Whitney Safety Platform Walker 800.255.3126 | wyeastmedical.com Operator’s Manual...
For assistance in troubleshooting, contact Wy’East Medical customer service or your Wy’East Medical representative. Include the following information with your request: 1. Serial and Model numbers of your Whitney Walker Safety Platform Walker. 2. Purchase date, if available. RETURNING ITEMS: To return parts or goods, contact Wy’East Medical for authorization.