PHASER User's Guide
PHASER is intended for use in appropriately equipped analytical laboratories. Solvents used
for analysis are flammable and toxic. Install the instrument in a very well ventilated room.
Provisions must be made for the storage of solvents in flame-proof cabinets and personnel
must wear protective clothing and eye protection at all times.
PHASER cannot be used on the GC where hydrogen (H
The PHASER controller must be powered from an earthed (grounded) mains outlet.
Power Requirements:
Maximum power consumption:
Both the PHASER controller and the host GC must be disconnected from the mains power
outlet until installation of the equipment is completed.
The PHASER controller power switch is located at the rear of the instrument. Leave enough
space behind the controller to ensure unrestricted access to the power switch.
Do not use power cords others than those supplied with the instrument as this may cause
fire or electric shock.
Ambient Conditions:

1.3 Operation Precautions and Warnings

Warnings:
There are potentially hazardous voltages present inside the PHASER control unit. To prevent
injuries and instrument damage, do not disassemble or modify it, or perform internal repairs.
When working with capillary column, it is important to wear the necessary protective
glasses to shield eyes from injury if the column breaks or shatters on impact.
The instrument is not explosion-proof. Do not use it in the vicinity of volatile solvents as
solvent vapor can ignite. If such a situation is unavoidable, ensure adequate ventilation in
the area during usage.
This instrument delivers GC injected compounds for direct absorption into the human body
through the nose. The use of dangerous compounds can result in serious injury. Carefully
check the safety aspects of the sample composition before using the instrument.
To prevent the risk of poisoning or death, do not use any liquid in the bubbling bottle other
than clean water. Always make sure the bubbling bottle is clean when using it. Always
clean the bubbling bottle after use.
Precautions and Warnings
AC Single Phase 220-240V, Frequency 50/60 Hz
230 VA
Temperature range: 18 - 40C
Humidity: 40 - 70%
) is employed as a carrier gas.
2
Issue 2.0
Page 1-2