BAS C3 Instruction Manual page 11

Cell stand
Table of Contents

Advertisement

Making Connections
Figure 3. Connection of C3 to BAS 100 Electrochemical Workstation
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Power Cord Connection
Push the socketed (female) end of the power cord into the port located in the back panel.
Before making this connection, make certain that the power on/off switch is in the off
position. This switch is labeled with a 0 and a 1. When 0 is pushed, the power is off and
when 1 is pushed, the power is on.
C3 to BAS 100 Series Electrochemical Analyzer
Two cables are required to connect the C3 to the BAS 100 series. Stirring and purging can
be automatically controlled by the BAS 100 series. The control cable is a 25-line ribbon
cable (ER-9515) between the REMOTE connector (25-pin D, male) on the back of the C3
and the CELL STAND connector (25-pin D, female) on the back of the BAS 100 series. See
Figure 3.
Connection of the cell lead cable is also shown in Figure 2. The cell lead cable has a
stainless steel LEMO connector at each end. One end is inserted into the CELL socket on
the back of the C3. The other end is inserted into the CELL socket in the back of the BAS
100 series. The ends are identical; therefore, it does not matter which end is connected to
the BAS 100 series or the C3.
Section 3. Installation
Cell Stand
Cell
CELL
GAS INLET
REMOTE
POWER ENTRY
MODULE
7