Amperometric detector

Advertisement

Table of contents

on page 3
User interface
on page 20

Specifications

Specifications are subject to change without notice.
Amperometric detector enclosure
Specification
Dimensions (W x D x H) Closed: 27.94 x 22.86 x 12.07 cm (11 x 9 x 4.75 in.); Open: 27.94 x 28.26 x 27 cm
Weight
Potentiostat
Specification
Dimensions (W x D x H)
Weight
Operating temperature
Operating humidity
Power requirements
Power output
Applied potential to flow cell ± 1.500 V
Measurement range
Offset
Output filter

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.
Details
(11 x 11.126 x 10.625 in.)
2.43 kg (5.35 lb)
Details
10 x 6.5 x 18 cm (4 x 2.5 x 7 in.)
0.9 kg (2 lb)
5 to 40 ºC (41 to 104 ºF)
80% maximum relative humidity for temperatures up to 31 ºC (88 ºF) decreasing
linearly to 50% at 40 ºC (104 ºF)
10 to 16 VDC unregulated at no-load, 100 mA/hour; Internal battery: NiCad,
standard 9 V rechargeable battery; Power supply: 12 VDC, 10 mA
0 to 1 VDC at I OUT port
Note: 1 VDC agrees with a full-scale reading in the selected range. For example, the I OUT port
signal is 0.5 V when the range setting is 10 nA and a flow cell current is 5 nA.
5, 10 and 20 nA range: ± 19.99 nA; 50, 100 and 200 nA range: ± 199.9 nA
± 199.9 nA
0.03, 0.08 or 0.3 Hz