7. Error reports
Error message
Description
Load cell defective
Operating error
Load cell defective
Overload
Adjustment with wrong adjustment
weight
Load cell defective
Balance not correctly levelled
The balance is in an environment
with strong air draught or
electromagnetic fields
Capacity of battery exhausted
8. Declaration of conformity
The current EC/EU Conformity declaration can be found online in:
www.kern-sohn.com/ce
EHA-BA-e-1910
Corrective measures
Replace load cell
Replace battery and readjust the balance
Replace load cell
Weighing only within the maximum load
Adjustment with correct adjustment weight
Replace load cell
Place the balance on a level, even
surface and level it using the four
adjustment feet in a way that the air
bubble of the spirit level is inside the
circle
Improve the environmental conditions
Insert new battery