Description: Sets the antenna delay value for calibration and system on-time-point. The
system will accept scientific notation values as shown below. The example below has a
delay of 96.3nsec
Syntax: antenna_delay [nt]
Command Example:
ATS-6501>antenna_delay 9.63E-8
Query:
ATS-6501>settings gps:antenna_delay
9.630000000000000e-08
Antenna Voltage:
Description: Sets the DC Bias voltage level on the Antenna connector. Default is Zero.
The system will give the user an error message with the acceptable values [0,5,12] if the
operator attempts to enter an invalid DC voltage.
Syntax: antenna_voltage [n]
Command Example:
ATS-6501>antenna_voltage 12
Query:
settings gps:antenna_voltage
ATS-6501>
0.0000000000000000e+00
5.0000000000000000e+00
1.2000000000000000e+01
Clear Alarms
Description: Clears any latched alarm events from the alarm list and front panel. This
allows the operator to acknowledge a system alarm and clear it from the status message
(status alarm) and clear the front panel alarm light. If any additional alarms/events occur
the new alarm will be added to the list (status alarm) and the front panel alarm light will
illuminate. If the condition the clear_alarms command was issued for clears and returns,
the alarm will be displayed in status alarm and the front panel alarm light will come on.
Syntax: clear_alarms
Command Example:
ATS-6501>clear_alarms
Latched alarms cleared
Config:
Description: Displays the current system configuration settings. These settings are done
and the factory and cannot be changed by the user. They are for information only. Short-
cut for show config.
0V
5VDC
12VDC
81
DOC06501_Revision F
ATS-6501 Users Guide