Table of Contents PSP Verification/ Adjustment Manual Table of Contents How to Use this Manual This manual describes how to verify and adjust the performance of PSP-603/ How to Use this Manual..........4 405/ 2010 Programmable Switching D.C. Power Supplies. Specification, page5, shows PSP electronic and mechanical specifications.
Specification PSP Verification/ Adjustment Manual Specification Front Panel Output Verification Adjustment - % key F key NORMAL key Decreases the + % key PSP-603 PSP-405 PSP-2010 Changes the Turns OFF the Output by a Increases the parameter in fine effect of + % and predefined value Voltage 0 ~ 60V...
Performance Verification PSP Verification/ Adjustment Manual Output Voltage Verification Performance Verification Here we verify the accuracy of Minimum and Maximum Output Voltage. Connection Overall Procedure Digital Multimeter Prepare the Equipment according to the following table. Verify a specification item and record the result (page8 to DC V Recording Table page14).
Performance Verification PSP Verification/ Adjustment Manual Output Current Verification Current Load Regulation Verification Here we verify the accuracy of Minimum and Maximum Output Current. Here we verify the amount of Output Current change when the load fluctuates. Connection Connection Digital Multimeter Connect the Digital Multimeter in series with the Electronic Load and PSP.
Performance Verification PSP Verification/ Adjustment Manual Verification steps Voltage Load Regulation Verification Connect the Multimeter and Electronic Load and turn ON the Here we verify the amount of Output Voltage change when the load Output. fluctuates. Set the Electronic Load level according to the Settings table on the previous page and turn ON the load.
Performance Verification PSP Verification/ Adjustment Manual Verification steps Ripple Voltage Verification Connect the Multimeter and Electronic Load and turn ON the Here we verify the amount of AC Ripple Voltage when the power source Output. fluctuates. Set the Electronic Load level according to the Settings table on the previous page and turn ON the load.
Performance Verification PSP Verification/ Adjustment Manual Verification steps Recording Table Connect the Multimeter and Electronic Load and turn ON the Output. PSP-603 PSP-405 PSP-2010 Set the Electronic Load level according to the Settings table on Multimeter Minimum Multimeter Pass/ Fail Note the previous page and turn ON the load.
Adjustment PSP Verification/ Adjustment Manual Opening the Case Adjustment Take off four screws, 3x6mm, on the side panels. Take off two screws, 3x6mm, on the top of the rear panel. Overall Procedure Prepare the Equipment according to the following table. Adjust the specifications according to your needs (page19 to 20).
Adjustment PSP Verification/ Adjustment Manual Output Voltage Adjustment Output Current Adjustment Here we adjust the accuracy of Minimum and Maximum Output Voltage. Here we adjust the accuracy of Minimum and Maximum Output Current. Connection Connection Digital Multimeter Digital Multimeter Adjustment steps Adjustment steps Connect the Multimeter.
Need help?
Do you have a question about the PSP 603 and is the answer not in the manual?
Questions and answers