2 Symbols 2. Symbols If the instructions are not followed properly, operation may WARNING lead to hazards for the product or the user/patient. Additional information, explanation of operation and performance. Serial number Catalogue number Date of manufacture Lot of manufacture Protective earth(ground) Type B applied part Fuse...
3.2 Safety instructions 1. Please read this manual before use. 2. The VacStation must be placed upright when being used. Laying on the side or upside down is prohibited, because these will cause damage to the machine or shorten its service life.
Page 7
HF Surgical Equipment in the hospital. Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the VacStation, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
4 Installing VacStation 4. Installing VacStation Step Diagrammatic Sketch Description Remove the top cover: Open four buckles (marked by blue arrows) and remove the top cover. Take out main filter assembly: Take out the sealing strip and filter elements in sequence. Then take out the main filter assembly enclosed by the dotted line.
Page 9
4 Installing VacStation Install the main engine box: Install the parts in order and fix four buckles. Installation of suction nozzle: Screw the nut, fix suction nozzle to the directional duct. Installation of directional duct: Plug the directional duct to the top cover.
Page 10
4 Installing VacStation Connecting power cord: Install one end of power cord into main engine housing, and the other end into a power outlet. Only the original power cord could be used. Connecting volume controller: Plug one end of the connecting wire into...
5 User Interface 5. User Interface Turn Power On Press power switch 4 to “ ” Volume up Shot press volume up key 3 to change air volume from 1 to 10. Volume down Shot press volume down key 2 to change air volume from 10 to 1.
6 Operating instructions 6. Operating instructions Do not block the air inlet artificially to avoid damage to the motor. After the filter is blocked, please replace the filter immediately to avoid damage to the motor. Before first use, be sure to remove the foam placed under the main filter. When using the buckle button, be sure to press the top cover part firmly, and then fix the buckle button, otherwise the buckle button will be damaged.
6 Operating instructions 6.3 Power off Turn off the power key on the external keyboard, and the window will display "off" before turning off. Do not use the device continually for more than 1 hour each time. Make sure no water accumulated on the suction nozzle.
7 Maintenance 7. Maintenance The components in the figure are: 1: Main engine housing 2: Main filter assembly 3: Secondary filter assembly 4: Primary filter cotton 5: Rubber pad 6: Top cover 7: Directional duct 8: Suction nozzle 7.1 Cleaning and disinfection 7.1.1 Forward For hygiene and sanitary safety purpose, the components (suction nozzle) must be cleaned and disinfected before each usage to prevent any contamination.
7 Maintenance should be reprocessed before sending back to the manufacturer for repair. 7.1.2 General recommendations The user is responsible for the sterility of the product for the first cycle and each further usage as well as for the usage of damaged or dirty instruments, where applicable after sterility.
7 Maintenance 7.2 Filter replacement 7.2.1 The frequency of filter replacement Components Frequency Primary filter cotton 1Month Secondary filter assembly 1~3 Month Main filter assembly 6~12 Month NOTE When the LED window displays "□" circularly, the filter must be replaced. 7.2.2 How to replace filters Release the four buckle buttons on the edge of the upper cover of the machine, and pick up the top cover part, primary filter cotton, secondary filter assembly and...
8 Trouble shooting 8. Trouble shooting When a problem or malfunction occurs, please check the machine with the table below before contacting the dealer to quickly eliminate common problems or malfunctions. If the problem or malfunction is not solved, please contact the dealer. Problem or Reasons Solutions...
9 Technical Data 9. Technical Data Manufacturer Changzhou Sifary Medical Technology Co.,Ltd Model VacStation Dimensions 65cm x 45 cm x 60 cm±1 cm(Package) Φ75mm×1500 mm Duct 21kg±10% Weight 220V AC ±10% for European Standard/110V AC ± Input 10% for American Standard...
10. EMC Tables Guidance and manufacturer’s declaration – electromagnetic emissions The VacStation is intended for use in the electromagnetic environment specified below. The customer or the user of the VacStation should assure that it is used in such an environment. Electromagnetic...
Page 20
10 EMC Tables Electrostatic discharge +/- 8 kV +/- 8 kV Floors should be wood, (ESD) IEC 61000-4-2 contact contact concrete or ceramic tile. If floors are covered with +/- 2 kV, +/- 4 +/- 2 kV, +/- 4 synthetic material, the kV, +/- 8 kV, kV, +/- 8 kV, relative humidity should...
Page 21
Guidance and manufacturer’s declaration – electromagnetic immunity The VacStation is intended for use in the electromagnetic environment specified below. The customer or the user of the VacStation should assure that it is used in such an environment. IEC 60601 test...
Page 22
The VacStation has been tested with the immunity test level in the below table and meet the related requirements of IEC 60601-1-2:2014. The customer and/or user should help keep a minimum distance between RF wireless communications equipment and the VacStation as recommended below.
Page 23
Use of VacStation adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, VacStation and the other equipment should be observed to verify that they are operating normally.
11 Statement 11. Statement Service Life The service life of VacStation series products is 12 months. Maintenance MANUFACTURE will provide circuit diagrams, component part lists, descriptions, calibration instructions to assist to SERVICE PERSONNEL in parts repair. Disposal The package should be recycled. Metal parts of the device are disposed as scrap metal.