2. Symbols used .................... 6 3. Before use ....................7 3.1Intended purpose ....................7 3.2Contraindications ....................7 4. Assemble and disassemble the MotorTurbo ......... 10 4.1Host connection ....................10 4.2Motor connection and disconnection ............10 4.3Handpiece connection and disconnection ..........11 4.4Insertion and removal of the bur ..............
Scope of MotorTurbo 1. Scope of MotorTurbo 1.1 Parts identification ① Host ② Motor ③ Dental Contra-angle Handpiece ④ Power cable ⑤ Power adapter ⑥ Motor disinfection front plug ⑦ Motor disinfection back plug ⑧ Spray nozzle ⑨ Needle 1.2 Components...
2 Symbols used 2. Symbols used General warning sign Caution Serial number Catalogue number Manufacturer Country of manufacture Batch code Authorized representative in the European Community Medical device CE marking 0297 Type B applied part Direct current Dispose of in accordance with the WEEE directive Keep dry Sterilizable in a steam sterilizer (autoclave) at...
(including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the MotorTurbo, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Page 8
3 Before Use If irregularities occur in the device during treatment, switch it off. Contact the agency. Please use the original power adapter. Connect to the power supply network with protective grounding. Do not dismount the motor or contra angle during operation, otherwise the contra angle and motor will be damaged.
Page 9
3 Before Use chunk is not fully released, otherwise, it will damage the handpiece. The clamping shaft length should not be too short, otherwise it will cause uneven bearing load and accelerate bearing abrasion. Do not use non Sifary accessories to avoid damage to the handpiece or other hazards.
4 Assemble and disassemble the MotorTurbo 4. Assemble and disassemble the MotorTurbo 4.1 Host connection Power Compressed Power Motor tail air and water connector switch wire connector Open the package, please make sure the product contains all accessories listed on the packing list, and place the host on a stable surface.
4 Assemble and disassemble the MotorTurbo 4.3 Handpiece connection and disconnection Motor connector Motor Handpiece O-ring Align the dental handpiece positioning block and the motor groove, insert the motor connector directly into the handpiece, and when you hear a "click" sound, the connection is completed.
Page 12
4 Assemble and disassemble the MotorTurbo Page 12 / 37...
5 Use interface 5. Use interface 5.1 Host operation panel There are 6 touch buttons on the host operation panel , their functions and uses are as follows: ① LED button: Short press to switch the motor LED luminance. ② Working mode button: Short press to switch working mode (M1-M4).
Use Interface according to their habits. The speed, handpiece transmission ratio, rotation direction and LED status of each working mode can be set independently, and the working mode can be automatically saved after each setting. It will enter the working mode of the last shutdown when it is turned on again.
Setting 6. Setting 6.1 Set mode parameter M1 mode parameter setting (connected to a handpiece with a transmission ratio of 1:5) Set the working mode to M1 Set the transmission ratio to Set rotation direction to Set LED light status to high-light rotation speed...
Page 16
Setting M3 mode parameter setting (connected to a handpiece with a transmission ratio of 16:1) Set the working mode to Set the transmission ratio to 16:1 Set rotation direction to Set LED light status to OFF Set rotation 600RPM (Rotation range is 100RPM - 2500RPM, and step value is 100RPM) M4 mode parameter setting (connected to a handpiece with a...
Operation 7.Operation Install product correctly according to the product installation steps; 2. Turn on the power switch on the back of the host and enter the standby state; 3. Set the spray air and cooling air input pressure, recommended pressure is 0.25-0.3MPa. 4.
Page 18
Operation When the operation completed, please release the foot switch of dental treatment unit first; then cut off the power; next disassemble the handpiece and motor; finally remove the power adapter and the handpiece four-hole connector of the dental treatment unit from the host.
Cleaning, disinfection and sterilization 8.Cleaning, disinfection and sterilization 8.1 Foreword For hygiene and sanitary safety purpose, the components (Motor, Motor disinfection front plug and Motor disinfection back plug) must be cleaned, disinfected and sterilized before each usage to prevent any contamination. This concerns the first use as well as the subsequent uses.
Page 20
Cleaning, disinfection and sterilization Dental Contra-angle Handpiece Only the components above can be autoclaved. Before first use and after each use, sterilize the above components. Reprocessing instructions Preparation at the point of use: Disconnect the motor from the host. Refer to "Chapter 4.2 - Motor connection and disconnection"...
Page 21
Cleaning, disinfection and sterilization each time. Do not rinse the motor with water, to prevent water from entering the motor. Do not submerge motor with any liquid. Cleaning for Dental Contra-angle Handpiece: Regarding cleaning/disinfection, rinsing and drying, it is to distinguish between manual and automated reprocessing methods.
Page 22
Cleaning, disinfection and sterilization recommended. Drying for motor: Manual Drying: Manual drying can be performed through lint free towel. Insufflate cavities of instruments by using sterile compressed air. Automated drying for Dental Contra-angle Handpiece: Drying of outside of instrument through drying cycle of washer/disinfector.
Page 23
Cleaning, disinfection and sterilization Do not lubricate the motor Packaging: Pack the instruments in an appropriate packaging material for sterilization. Check the validity period of pouch given by the manufacturer to determine the shelf life. Use pouches which resist to a temperature up to 141℃ and in accordance with EN ISO 11607.
Page 24
Cleaning, disinfection and sterilization to ensure that the processing, as actually performed using equipment, materials and personnel in the processing facility, achieves the desired result. This requires verification and/or validation and routine monitoring of the process. Likewise, any deviation by the processor from the instructions provided should be properly evaluated for effectiveness and potential adverse consequences.
Cleaning, disinfection and sterilization Disinfection components Disinfection components Host Power cable Adapter Wipe all the surfaces with a cloth lightly moistened with Ethanol for Disinfection (Ethanol 70 to 80 vol%) at least 2min, repeat for 5 times. Do not use disinfectants other than ethanol for disinfection (Ethanol 70 to 80 vol%).
Error warnings 9.Error warnings When input source pressure is more than 0.6MPa, the fault light will be on. At this time, the host will not work while operating the control foot switch. Adjust the input pressure to the range of 0.25-0.5MPa. If the fault light is still on, please contact your local dealer.
Troubleshooting 10.Troubleshooting When trouble is found, check the following points before contacting your distributor. If none of these are applicable or the trouble is not remedied even after action has been taken, the product may have failed. Contact your distributor. Problem Cause Solution...
Page 28
Troubleshooting Unauthorized modification of the MotorTurbo may cause hazards: The motor can’t reverse. Water or air leakage from motor. The host can’t control the motor. Circuit board damage Page 28 / 37...
Technical data 11.Technical data Manufacturer Changzhou Sifary Medical Technology Co., Ltd Model MotorTurbo Power adapter input AC 100~240V, 50/60Hz, ±1Hz 1.5A Power adapter output 2.5A T3.15AL 250V Power adapter fuse Main unit input 2.5A 5NL/min≤Q≤40NL/min Cooling air (Q) (0.25MPa-0.5MPa) Motor Spray air supply ≥1.5NL/min(0.25MPa)...
Page 30
Technical data Use: in enclosed spaces Ambient temperature: 5°C ~ 40 °C Operation conditions Relative humidity: <80% Operating altitude < 3000m above sea level Ambient temperature: -20 °C ~ +55 °C Transport and storage Relative humidity: 20% ~ 80 % conditions Atmospheric pressure: 70kPa ~ 106kPa Page 30 / 37...
Guidance and manufacturer’s declaration - electromagnetic emissions The MotorTurbo is intended for use in the electromagnetic environment specified below. The customer or the user of the MotorTurbo should assure that it is used in such an environment. Complian Electromagnetic Emissions test...
Page 33
Guidance and manufacturer’s declaration – electromagnetic immunity The MotorTurbo is intended for use in the electromagnetic environment specified below. The customer or the user of the MotorTurbo should assure that it is used in such an environment. Electromagnetic IEC 60601 Compliance Immunity test...
Page 34
When they are used in close proximity to medical equipment and/or systems, the medical equipment and/or systems’ basic safety and essential performance may be affected. The MotorTurbo has been tested with the immunity test level in the below table and meet the related requirements of IEC 60601-1-2:2020.
Page 35
Guidance and manufacturer’s declaration –Proximity magnetic fields The MotorTurbo is intended for use in the electromagnetic environment specified below. The customer or the user of the MotorTurbo should assure that it is used in such an environment. Proximity IEC 61000-4-39 Complian...
Dental Contra-angle Handpiece is 6 months starting from the day of delivery to the customer 2. MotorTurbo should be repaired by the equipment technology department of Changzhou Sifary Medical Technology Co., Ltd. or maintenance service partners authorized by Changzhou Sifary Medical Technology Co., Ltd.
14.Statement Shelf Life The shelf life of MotorTurbo series products is 5 years. To ensure continued safe use of MotorTurbo. It is recommended that the equipment (motor and host) be checked and repaired at the dealer once a year. Maintenance...