1.2 Check Order Check, the contents of the shipment against the purchase order. Verify that the CP-10 is received as ordered. Each CP-10 is labeled with its target gas. If there are accessories on the order, ascertain that they are present. Check the contents of calibration kits.
See Section 4.0 and 5.0 for operational and maintenance flow charts. Three alarm points are preprogrammed into the CP-10. At each alarm point, an LED on the front panel is activated. There are 4, 10 Amp relay contacts at each alarm point, plus a fault relay. See Section 3.2 for wiring information.
4-20 mA output. See Section 3.1.2 Sensor Terminal J18 Sensor/Transmitter connection, See Section 3.2 Data Terminal J19 RS-485 input/output Potentiometer 2, 3 & 4 Not used in CP-10 Do Not Adjust Sensor/Transmitter RS-485 Input/Output Connection Terminal J18 Terminal J19 POT 2...
ENMET Corporation 3.0 Installation of the CP-10 The CP-10 is supplied with a strain relief for a power line cord. Use this fitting or connect a conduit fitting when supplying power to the unit. : This control panel is N rated for hazardous locations.
ARNING The compliance of the installation to appropriate codes is not ENMET’s responsibility. The CP-10 should be powered through appropriately sized circuit breakers. See Section 6.0 Technical Data. 3.1.2 Power Supply The input power can vary from 100 to 240 V , 50/60 Hz.
If you have a question involving the location of a unit or sensor, please contact your distributor or ENMET personnel. A technician will analyze the question and recommend a location.
It is recommended that power for auxiliary equipment be supplied from an independent power source separate form the CP-10. Use the existing hole in the enclosure for wire to enter and exit and use appropriate cable fittings. Wiring should...
4.0 Operation When the CP-10 is installed as described in Section 3, and in clean air, the POWER green LED is on, the display is lit and the information on the display is measurement of the target detected by the CP-10. The red alarm and fault LEDs are not lit.
4.2 Normal Display Mode When the CP-10 is installed as described in section 3, the POWER green LED is on, the display is lit and the measurement designator: ppm, LEL or % is displayed by the CP-10. The red alarm and fault LEDs are not lit.
Press and hold the M button for 2 – 4 seconds to enter the Maintenance Menu The Power/Fault LED will flash Green – Red to indicate the CP-10 is in Maintenance Mode To exit the maintenance Menu and return to the...
Press Select button to set low end of scale: Menu Menu The Select button locks underscored digit and moves to next digit The Menu button changes digit indicated by underscore cursor When last digit is entered CP-10 will advance to full scale set up Select Select ² ²...
Press Select button to set low end of scale: Menu Menu The Select button locks underscored digit and moves to next digit The Menu button changes digit indicated by underscore cursor When last digit is entered CP-10 will advance to full scale set up Select Select 30.0 ²...
5.2.2 Set Up 4 and 20 Scale The CP-10 4 mA low scale (the low end of range) and 20 mA full scale (the high end of range) is set at the factory. The mA Trim is for the adjustment of miner differences that may occur between the CP-10 control and the S/T display.
Software revision may cause variations of display output. 5.2.4 mA Span Set The CP-10 4-20mA span range is set at the factory, normally to the full scale of the measurement and can be changed within limits. To change the span range: Enter the maintenance menu as shown in Figure 6 CP-10 Maintenance Menu flow chart.
CP-10 ENMET Corporation 6.0 Technical Data and Specifications The CP-10 technical data and specifications: Electrical Power 15 Amp fused branch circuit 100-240 V 0.45A, 50/60 Hz 0.6A, 24V Storage and Transport Temperature: -20° to +60°C (-4° to +140°F) preferred 0° to +20°C (32° to 68°F)
ENMET shall not be liable for any loss or damage caused by the improper use of the product. The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise through the use by the purchaser or others of this equipment.