Olympus OEV262H Instructions For Use Manual

Olympus OEV262H Instructions For Use Manual

High definition lcd monitor

Advertisement

Instructions for Use
HIGH DEFINITION LCD MONITOR
OEV262H
USA: CAUTION: Federal law restricts this device to sale by or on the order of a physician.
Table of Contents
Labels and Symbols
Important Information -
Please Read Before Use
Features
Location and Function
of Parts and Controls
Installation and
Connection
Troubleshooting
Appendix
1
2
3
4
5
6

Advertisement

Table of Contents
loading

Summary of Contents for Olympus OEV262H

  • Page 1 Features Location and Function of Parts and Controls Installation and Connection Instructions for Use Inspection Setting HIGH DEFINITION LCD MONITOR OEV262H Troubleshooting Appendix USA: CAUTION: Federal law restricts this device to sale by or on the order of a physician.
  • Page 2 OEV262H...
  • Page 3: Table Of Contents

    System chart ............38 On Repacking ............7 EVIS video system chart ........39 On Fan Error ............7 Camera control unit system chart .....40 Connection cable for OEV262H .......41 On Moisture Condensation ........7 Specifications ............42 Signal Words ............7 Dimensions ............47 Warnings and Cautions ........
  • Page 4: Labels And Symbols

    Safety-related labels and symbols are attached to the Rear Power switch. instrument at the locations shown below. If labels or Press to turn the monitor symbols are missing or illegible, contact Olympus. High definition LCD monitor Rear Power switch. Press to turn the monitor off.
  • Page 5: Important Information - Please Read Before Use

    Chapter 6, contact Olympus. This monitor is to be repaired by Olympus technicians only. Light source: Olympus is not liable for any injury or damage which The light source provides light and electrical signals to occurs as a result of repairs attempted by non-Olympus the endoscope.
  • Page 6: On Installation

    • The screen and the cabinet become warm during assets or stable picture, or for emergencies, we strongly operation. This is not a malfunction. recommend you use more than one unit or prepare a spare unit. Important Information — Please Read Before Use OEV262H...
  • Page 7: On Repacking

    (by air-conditioning, for example), stop operating it and contact Olympus. moisture may condense on the surface or inside of the – Do not prepare, inspect or use this monitor with wet unit, or create a mist residue inside the protection plate.
  • Page 8: Inspection

    • Do not use this monitor beyond the rated outlets and confirmation, patient injury may result. wiring, otherwise it may result in fire or an electrical • It is recommended that only Olympus high-frequency shock. electrosurgical equipment will be used with this unit.
  • Page 9 To prevent risk of electric shock Class A, Group 1) or fire hazard due to overheating, ensure that curtains 6.Model OEV262H is a monitor intended for use in a and any other materials do not obstruct the ventilation. medical environment to display pictures from cameras •...
  • Page 10 • The response speed and brightness of liquid crystal wiping material. The screen and the panels will be varies with ambient temperatures. scratched. • Some video images may appear blurred on the screen. Important Information — Please Read Before Use OEV262H...
  • Page 11 (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. For the customers in Canada CAN ICES-3 (A)/NMB-3(A) Important Information — Please Read Before Use OEV262H...
  • Page 12: Features

    Input connectors COMPOSITE input connector × 1 Y/C input connector × 1 HD15 input connector (RGB/COMPONENT compatible) × 1 SDI input connector (3G/HD/SD compatible) × 2 DVI input connector (DVI-D/ANALOG RGB) × 2 1.1 Features OEV262H...
  • Page 13 This reduces power consumption. Select language display You can select the display language from English, French, German, Spanish, Italian, and Japanese. Control Lock function You can use the Control Lock function to prevent malfunction. 1.1 Features OEV262H...
  • Page 14: Location And Function Of Parts And Controls

    PIP/POP: For displaying the multi-image display or Press to select the items and setting values. switching the multi-image display mode. SWAP: For swapping between the main display and the sub display. 2.1 Front Panel OEV262H...
  • Page 15 PORT A: Press to monitor the signal assigned to PORT A. PORT B: Press to monitor the signal assigned to PORT B. AUX IN: Press to monitor the signal through the AUX IN input connector. 2.1 Front Panel OEV262H...
  • Page 16: Input Signals And Adjustable/Setting Items

    *4 Sync Detect can be switched but is not applied when the HD15/DVI or analog PC signal is input. *5 The setting value can be changed but is not applied to the screen when the DVI or analog PC signal is input. 2.2 Input Signals and Adjustable/Setting Items OEV262H...
  • Page 17: Rear Panel

    OUT connector, the monitor may affect the operation of the connected equipment. 2.3 Rear Panel OEV262H...
  • Page 18 DVI input/output connector (DVI-I) [1] IN connector, [2] IN connector Input DVI Rev.1.0 applicable digital signal, analog RGB signal (0.7 Vp-p, positive), or sync signal (HV SYNC). [1] OUT connector, [2] OUT connector Active through output connector. 2.3 Rear Panel OEV262H...
  • Page 19: Installation And Connection

    • If a trolley other than the mobile workstation (WM-NP2, WM-DP2, WM-NP1, WM-WP1, or WM-DP1) is used, confirm that the trolley can withstand the weight of the equipment installed on it. 3.1 Installing the monitor / 3.2 Connecting the AC Power Cord OEV262H...
  • Page 20: When The Mobile Workstation (Wm-Np2, Wm-Dp2, Wm-Np1, Wm-Wp1, And Wm-Dp1) Is Used

    When the mobile workstation (WM-NP2, WM-DP2, WM-NP1, WM-WP1, and WM-DP1) is used AC IN OEV262H Mobile workstation Power socket Power cord provided with the mobile workstation to the wall mains outlet Power cord of the mobile workstation Place the mobile workstation on level surface. Lock the caster brakes by pushing them down.
  • Page 21: To Use The Ac In Socket On The Monitor

    AC IN socket AC power cord Plug holder for the supplied AC power cord (supplied) Slide the plug holder for the supplied AC power cord over the cord until it locks. 3.2 Connecting the AC Power Cord OEV262H...
  • Page 22: To Use The Optional Ac Adapter

    AC power cord to the AC power cord. To use the AC adapter: Insert the DC IN connector into the DC 5V/24V IN connector until it locks. Otherwise the cable may fall out and an image may disappear. 3.2 Connecting the AC Power Cord OEV262H...
  • Page 23: About The Connecting Cables

    3.3 About the Connecting Cables For details about the connecting cables, refer to “Connection cable for OEV262H” on page 41. 3.3 About the Connecting Cables OEV262H...
  • Page 24: Inspection

    Olympus. its instruction manual and make sure that the noise does Using the LCD monitor (OEV262H) while an not affect the observation of surgical procedures. If irregularity is observed does not only result in...
  • Page 25: Setting

    “Off” by referring the following steps 4 to German, Spanish, Italian, Japanese) for displaying the menu and other on-screen displays. NOTE OEV262H MENU If the Control Lock has been turned on, all items are Color Tone Adjustment displayed in gray. To change any of the items, turn the Screen Control Control Lock to OFF first.
  • Page 26: Using The Menu

    Press the b/B button to select a language. 5.2 Using the Menu The menu changes to the selected language. The unit is equipped with an on-screen menu for making OEV262H System Configuration various adjustments and settings such as picture control, Control Power On input setting, setting change, etc.
  • Page 27: Adjustment Using The Menus

    Press the v/V button to select a menu. 5.3 Adjustment Using The selected menu appears and setting items of the selected tab are displayed. the Menus OEV262H Color Tone Adjustment SDI 1 SDI 2 DVI 1 DVI 2 Items Color Mode...
  • Page 28: Adjusting And Changing The Settings

    You need to use the measurement instrument to adjust the white balance. Control Lock Recommended: Konica Minolta color analyzer CA-310 OSD Setting Power On Setting Power Save OEV262H Remote Color Tone Adjustment Custom Button SDI 1 SDI 2 DVI 1...
  • Page 29 The Screen Control menu is used to set the image display setting for each input. PIP/POP menu The PIP/POP menu is used to set the display mode for the multi-image display and for each input. OEV262H Screen Control Example “Clipping Size” menu SDI 1 SDI 2...
  • Page 30 The System Configuration menu is used to set the AUX IN settings. Control Lock, OSD Setting, Power On Setting, Power Save, Remote settings, Custom Button setting, and Example “Clone Out” menu Panel Display setting. OEV262H Example “Control Lock” menu Input/Output Configuration Port A Port B Clone Out...
  • Page 31 • Contrast connector for more than 1 minute. Panel Display Backlight Adjusts the brightness of the display. A higher setting increases the brightness of the display, and a lower setting darkens the display. 5.3 Adjustment Using the Menus OEV262H...
  • Page 32 PORT A/PORT B HD Sets the HD clipping size for each Clipping Size port when displaying multi-image Example “Default Preset” menu display. Select from among “Normal,” “4:3,” “5:4,” “Square1,” OEV262H “Square2,” “Square3,” “Square4,” Initial Setup “Square5.” Default Port A Port B...
  • Page 33 Composite Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip Not skip 5.3 Adjustment Using the Menus OEV262H...
  • Page 34 Flip Flip Contrast Contrast Contrast Contrast Contrast Custom Scan Scan Button Button2 Brightness Brightness Brightness Brightness Brightness Brightness Brightness Brightness Size Size Backlight (Energy Saving Mode OFF) Panel Display Backlight (Energy Saving Mode ON) 5.3 Adjustment Using the Menus OEV262H...
  • Page 35 VISERA Video system center OTV-S7V EVIS EXERA II Video system center CV-180 Preset I (for NTSC model) VISERA Pro Video system center OTV-S7Pro Preset J (for PAL model) It does not use Preset A, F 5.3 Adjustment Using the Menus OEV262H...
  • Page 36: Troubleshooting Guide

    Troubles or failures other than those listed in the following table need repair. As repair performed by WARNING persons who are not qualified by Olympus could cause Never use the monitor if an abnormality is suspected. patient or user injury and/or equipment damage, be sure Damage or irregularity in the instrument may to contact Olympus for repair.
  • Page 37: Returning The Monitor For Repair

    6.2 Returning the monitor for repair CAUTION Olympus is not liable for any injury or damage that occurs as a result of repairs attempted by non-Olympus personnel. When returning the monitor for repair, contact Olympus. With the monitor, include a description of the...
  • Page 38: Appendix

    New products released after the introduction of this monitor may also be compatible for use in combination with it. For further details, contact Olympus. WARNING If combinations of equipment other than those shown below are used, the full responsibility is assumed by the medical treatment facility.
  • Page 39: Evis Video System Chart

    Endoscope position (see page 41) detecting unit UPD-3 SDI cable (OEP-5) MAJ-1464/1912/1951 BNC cable (OEP-4) MB-677 High definition LCD monitor OEV262H Color video printer OEP-5, OEP-4 SDI cable MAJ-1464/1912/1951 Mobile workstation WM-NP2, WM-DP2, WM-NP1, Keyboard Video recorder WM-WP1, WM-DP1...
  • Page 40: Camera Control Unit System Chart

    Camera control unit system chart Combination of connection cables High definition LCD monitor SDI cable (see page 41) OEV262H MAJ-1464/1912/1951 Video recorder IMH-20, IMH-10 SDI cable MAJ-1464/1912/1951 DVI cable Video system center MAJ-1945 OTV-S190, OTV-S7Pro, OTV-S7V Camera head Wireless transmitter...
  • Page 41: Connection Cable For Oev262H

    Connection cable for OEV262H CV-290, CV-260SL, CV-190, CV-180, CV-165, CV-170, CV-150, EU-ME2, EU-ME2 PLUS PREMIER, EU-ME2 PLUS, EU-ME1 SDI cable DVI cable Monitor cable MAJ-1160: MAJ-1462: Video system center / MAJ-1912: MAJ-1228: Endoscopic MAJ-1464: 2.5m MAJ-1945: MAJ-1584: 1.5m MAJ-1592: ultrasound center...
  • Page 42: Specifications

    (Sync On Green, connector 0.3 Vp-p sync negative) Component: 0.7 Vp-p ± 3 dB (75% chrominance standard color bar signal) • External synchronized input BNC type (×1) 0.3 Vp-p to 4.0 Vp-p ± bipolarity ternary or negative polarity binary Specifications OEV262H...
  • Page 43 IEC 60601-1-2: 2001 IEC 60601-1-2: 2007 This instrument complies Power LCD monitor with the EMC standard (OEV262H) for medical electrical AC 100 V-240 V, 50/60 equipment, edition 2 Hz, 0.96 A-0.47 A (IEC 60601-1-2: 2001) DC IN: 24 V 3.6 A 5 V and edition 3 (IEC 0.03 A (Supplied from...
  • Page 44 DVI 1 SDI 2 (SDI) DVI 2 3G/HD/SD-SDI DVI-D 480/59.94i – 480/60p – – 575/50i – Pin number Signal 576/50p – – 720/59.94p 720/60p 720/50p 1080/59.94i 1080/60i 1080/50i 1080/59.94p 1080/60p 1080/50p The frame rate is also compatible with 1/1.001. Specifications OEV262H...
  • Page 45 1920 × 1200 60 Hz 154.125 74.099 59.999 Positive Negative 1280 × 1024 60 Hz 91.000 63.194 59.957 Positive Negative 1280 × 768 50 Hz 65.125 39.518 49.959 Negative Positive 1280 × 768 60 Hz 80.125 47.693 59.992 Negative Positive Specifications OEV262H...
  • Page 46 Range of DVI input signal (available to 1920 × 1080/ 60 Hz) Vertical frequency: 50.0 Hz to 85.1 Hz Horizontal frequency: 31.5 kHz to 77.0 kHz Dot clock: 25.175 MHz to 162.000 MHz Picture size, phase: automatic discrimination by the DE (Data Enable) signal Specifications OEV262H...
  • Page 47: Dimensions

    264.8 (10 100 (4) 100 (4) 100 (4) Unit: mm (inches) Mass Approx. 1.2 kg (2 lb 10 oz) * Length of M4 screws (4) Monitor cabinet Attached object M4 screw ) to 9 ( Unit: mm (inches) Dimensions OEV262H...
  • Page 48: Emc Information

    OEV262H. Guidance and manufacturer’s declaration-electromagnetic emissions The OEV262H is intended for use in the electromagnetic environment specified below. The customer or the user of the OEV262H should assure that it is used in such an environment. Emissions test Compliance...
  • Page 49 Guidance and manufacturer’s declaration - electromagnetic immunity The OEV262H is intended for use in the electromagnetic environment specified below. The customer or the user of the OEV262H should assure that it is used in such an environment. Immunity test IEC 60601 test...
  • Page 50 Guidance and manufacturer’s declaration - electromagnetic immunity The OEV262H is intended for use in the electromagnetic environment specified below. The customer or the user of the OEV262H should assure that it is used in such an environment. Portable and mobile RF communications equipment should be used no closer to any part of this model, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
  • Page 51 Recommended separation distances between portable and mobile RF communications equipment and the OEV262H The OEV262H is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the OEV262H can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the OEV262H as recommended below, according to the maximum output power of the communications equipment.
  • Page 52: Maintenance

    Maintenance Olympus recommends having Olympus endoscopic equipment (endoscopes, reprocessors, light sources etc.) inspected by an Olympus certified service technician at least once a year, to maintain the proper function of the equipment. OEV262H...
  • Page 53 OEV262H...
  • Page 54 OEV262H...
  • Page 55 ©2014 OLYMPUS MEDICAL SYSTEMS CORP. All rights reserved. No part of this publication may be reproduced or distributed without the express written permission of OLYMPUS MEDICAL SYSTEMS CORP. OLYMPUS is a registered trademark of OLYMPUS CORPORATION. The product names, logos and company names in this instruction manual are generally trademarks or registered trademarks of each company.
  • Page 56 RA0072 01 Printed in Taiwan 4-446-574-13(1)

Table of Contents