Warning messages display important information about the unit. The
following section describes the protocol for each message. Some
warnings automatically reset, while others require user intervention.
Visual warnings will always be accompanied with audio beeping.
Warning – Power Supply Failure
POWER SUPPLY FAILURE
Warning – Lamp Temperature
LAMP BOX TEMPERATURE
HIGH, ENSURE PROPER
Rev D
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
WARNING!
REFER TO
USER MANUAL
WARNING!
VENTILATION
This warning indicates a
power supply failure.
lamp will be powered down
and remain off. Cycle the
AC power.
If the problem
persists, call for technical
help.
This
warning
ventilation problem which
could be a warm ambient
environment,
vents, or a fan failure.
Assure proper
(clean vents and clearance).
The warning will clear if the
internal temperature of the
unit decreases below the
warning temperature. If the
message persists, call for
technical help.
The
signals
a
obstructed
ventilation
16