setting, the display will switch to show "bASIC" above the phrase "dEC".
This signals that the DLAF930SD is ready for your input in Step 4 of the
setup sequence.
4. Choose a decimal point or comma to represent the decimal division between
integers and fractions (for example, American-style 20.88 vs. European-style
20,88). Once you have set the date and time, the display will show the word
"bASIC" above the phrase "dEC". Press the SET M button to make "bASIC"
(American style) or "Euro" appear in the upper display, as desired. Press the
REC/ENTER button to store the selection as the default.
5. Enable or disable auto power off. Once the format of decimal point divisions
has been set and stored, the display will show the word "yES' over the term
"PoFF". Press the SET M button until the desired automatic power off
management scheme ("yES" for enable; "no" for disable) is displayed. Press the
REC/ENTER button to store the selection. If enabled, the power off function shuts
off the DLAF930SD after a period of inactivity of ten minutes.
6. Enable or disable the beeper. Once the auto power off function has been
enabled or disabled, the display will show the word "yES" over the word "bEEP".
Press the SET M button until the desired setting ("yES" or "no") is displayed,
and then press the REC/ENTER button to store the selection as the default.
7. Select the thermocouple type. Once the beeper has been enabled or disabled,
the lower part of the display will show the term "tYPE". Press the SET M or
LOGGER
button until your thermocouple's type (K or J) appears on the display.
Then press the REC/ENTER button to store the selection.
8. Select the temperature unit. Once the thermocouple type has been selected
and stored, the lower part of the display will show the term "t-CF". Press the SET
M or LOGGER
Celsius") is displayed. Then press the REC/ENTER button to store the selection
as the default.
9. Set the datalogging sampling time. Once the default temperature unit has been
selected and stored, the display will show a value above the letters "SP-t". Press
the SET M or LOGGER
desired sampling time appears above "SP-t". The options are 0, 1, 2, 5, 10, 30,
60, 120, 300, 600, 1800 and 3600 seconds (0 seconds to 1 hour).
10. Format the SD card. Once the sampling time has been set and stored, the term
"Sd-F" will appear in the lower half of the display. Press the REC/ENTER button.
Pressing the REC/ENTER button causes the word "no" or "yES" to appear over the
term "Sd-F". Press the M button to make a selection. Choose "yES" whenever a
new SD card is being used, or when a used card is being repurposed (from use
with another SD card instrument or a camera, for example) and all data on it is
to be erased. Choose "no" to preserve any data on a card previously used with
this instrument.
button until the unit you prefer ("F" for Fahrenheit or "C" for
button to decrease or increase the value until the
7