DigiTech QM-1549 Calibration Instructions
DigiTech QM-1549 Calibration Instructions

DigiTech QM-1549 Calibration Instructions

Advertisement

Quick Links

Calibration Operation Instruction
PICTURE INSTRUCTION
adjust DCV potentiometer
VR1
product name:
product model:
working procedure:
OPERATION STEPS
Debugging indication:           
  Check datum mark:input DC100mV,check and tune
potentiometer VR1 
  Debugging step:set tap position in DCV,input 100.0mV
by use standard DCV current source,check and tune
potentiometer VR1,make display range of LCD is
99.9mV~100.1mV
Test indication:
   
Measuring range Calibrating spot Normal display
range
400mV
0mV
0.0~0.1mV  
400mV
100.0mV
99.5~100.5mV  
4V
1.000V
0.990~1.010V  
40V
10.00V
9.90~10.10V  
400V
100.0V
99.0~101V  
1000V
600V
593~607V
Digital Multimeter
QM-1549
DCV

Advertisement

loading

Summary of Contents for DigiTech QM-1549

  • Page 1 Calibration Operation Instruction Digital Multimeter product name: QM-1549 product model: working procedure: PICTURE INSTRUCTION OPERATION STEPS Debugging indication:              Check datum mark:input DC100mV,check and tune potentiometer VR1    Debugging step:set tap position in DCV,input 100.0mV...
  • Page 2 Calibration Operation Instruction Digital Multimeter product name: product model: QM-1549 working procedure: PICTURE INSTRUCTION OPERATION STEPS 1、adjust ACA0 Set tap position in AC10A,adjust potentiometer VR1,let LCD display 0.000-0.001A 2,Check datum mark: Set tap position in ACV,input 100mV(60Hz) by use standard ACV current source ,let LCD display 0.999V~1.001V...