10 Displayinformation / Error messages

Date over display scope
F----1
A/D-switch disorder
F----2
Function key disorder
F----3
Sensor signal disorder / Load Cell defect
F----5
Lower zero position of sensor signal
F----L
Weighing capacity is exceeded
F----H
Calibration error, instability or not at zero point during calibrating
C----F
Calibration error, items on the pan or at higher zero point during calibrating
C----H
Storage calibration data error, re-calibration is necessary
E-----
Zero point of the scale disorder / mechanical defect of the scale
------
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
11 Configuration & User Menu
The settings of the balance can be changed in the setting menu. This way, the balance can be adjusted to
individual weighing requirements. By the factory the user menu has been set in a way that normally no
more changes must be made, only at special conditions of use.
Overview:
In the off state press the [CAL] button and keep it pressed, press the [ON/OFF] button turn on the balance.
Now the balance is accessed to the setting menu. Using the [CAL] button the individual menu items (C1-
C3) can be selected one after the other. Press the [TARE] button to change the setting value. After all
points have been configured, press [CAL] button returns the balance automatically into weighing mode.
a) C1 = sensitivity setting - values: 0 1 2 3 4
The higher the value, the lower the sensitivity and higher stability.
Filter 0: Setting for dispensing.
Filter 1/2: Sensitive and fast, very quiet using location.
Filter 3/4: Robust but slow, busy using location.
Default setting is 2 or 1.
b) C2 = filtering strength setting - values: 0 1 2 3
The higher the value, the lower the sensitivity and higher stability.
Filter 0: Setting for dispensing.
Filter 1/2: Sensitive and fast, very quiet using location.
Filter 3: Robust but slow, busy using location.
Default setting is 2 or 1.
c) C3 = serial port setting - transmission speed (RS-232)
2(600), 3(1200), 4(2400), 5(4800), 6(9600) - Default setting is 6.
G&G GmbH | Displayinformation / Error messages