Download; Show Data - Lafayette 01163 User Manual

Manual muscle test system
Table of Contents

Advertisement

Model 01163
Download Function (continued):
To download:
1.
Plug in the USB download device to the host computer
2.
Run the download application software.
During download, the MMT will
attempt to establish communication with
the computer. The computer must have
the download software running and the
USB download peripheral connected for
communication to be established.
3.
Point the Infrared port in the bottom of the MMT at the USB download device.
Since the connection is infrared, the line of sight between the computer and MMT must
be free from obstructions. The infrared transmitter on the bottom of the MMT must be
pointed directly at the USB download device. The MMT can not be more than 12 feet from
the download device. If the MMT is more than 12 feet away, errors may result in the data
transmission.
4.
Enter the Main Menu by pressing the MENU button
5.
Press ENTER to begin download.
If a connection can not be established, an error is given along with two audible beeps. An
option menu is then given which allows (1) a reattempt of communication or (2) canceling
the operation. The download device receives the data and transfers it to the computer
software. The software then checks the data for validity and reports the checksum back to the
MMT. Once the MMT receives the correct checksum, the data registers are cleared. If the
correct checksum is not received , the error message is given and the data is not cleared.

Show Data:

The Show Data function allows the
stored data to be display and reviewed.
To review stored data:
1.
Enter the Main Menu by pressing
the MENU button.
2.
Press the SCROLL button one (1) time to move the cursor to SHOW DATA.
3.
Press ENTER to display the SHOW DATA menu.
4.
Press the SCROLL button one (1) time to move the cursor to SHOW DATA.
5.
Press ENTER to show the data.
6.
Press the SCROLL button to step through the data registers.
7.
Press the RESET button to exit to the Main Measurement Screen.
Note:
Only the Peak Force (in lbs. or kg) and the peak time are stored in memory. Empty
data registers are shown as "7. KL G . S". The SHOW DATA function can also be
used to determine how many empty registers are available.