2.2.1 Model 8032-27 and Load Cell Wiring

Connect the "common" VW- conductor from the load cell to the 8032-27 by lifting up the
orange tab located on the opposite side of the six black conductors, inserting the common
conductor fully into the 8032-27 (Figure 11), and then pushing down on the orange tab
until it snaps into place. Refer to Table 3 to identify which conductor carries the common
VW- signal.
10 pin Bendix
Function
PT06A-12-10P
H
Common
*White's black and Green wires are switched on GEOKON 3-gauge VW load cells prior to serial number 3313.
The following wiring chart details the connections between the load cell and 8032-27
with the terminal board:
Table 4 - Standard VW Load Cell Wiring When Using 8032-27
Figure 11 - Model 8032-27 Jumper Wire Assembly
3 Gauge VW Load
Cell, Purple Cable
White's Black*
Table 3 - Common Conductor Chart
Terminal
Vibrating Wire
Board
with Thermistor
1H
1L
2H
2L
S1
Shield Drain Wire
3H
3L
4H
4L
S2
5H
5L
6H
6L
S3
11H
11L
12H
12L
S6
4 Gauge VW Load
Cell, Purple Cable
Green
VW Sensor #1
8032-27
Thermistor
Thermistor
VW Sensor #2
8032-27
-
-
-
VW Sensor #3
8032-27
-
-
-
VW Sensor #6
8032-7
-
-
-
9