Resetting To Factory Settings; Error Messages And Error Clearance - Weller WTHA 1 Supplementary Operating Instructions

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 10

Resetting to factory settings

When switching on: Exit, UP and DOWN until a signal sounds

Error messages and error clearance

Message/symptom
No display function (display
OFF)
OFF
Tool remains cold
Station is not operating as it
usually does
Settings cannot be changed
Insufficient quantity of hot air
Zero Smog is not running
Err 1
Err 2
+
Possible cause
No mains supply voltage
Station is on standby or in OFF mode
Station is on standby or in OFF mode
Mains voltage fault
Parameters set incorrectly
Station locked
Hose damaged, kinked or twisted
No mains supply voltage
No signal detected
Incorrect mains voltage
Heating cartridge defective
3
sec
Remedial measures
Turn on mains power switch
Check mains supply voltage
Check device fuse
Reactive the soldering tool using the UP
or DOWN buttons
Press the finger-operated switch or
foot-operated switch
Check the mains power supply
Reset the station to the factory settings
Unlocking the station
Disconnect the device from the mains
voltage
Check the hose
Have the device repaired by a specialist
Check the mains power supply
Check the interface wiring
Check the interface settings
Check mains supply voltage
Have the heating cartridge replaced a
specialist
GB