Maintenance and cleaning of the device
- In case of damaged case, this should be replaced and no longer used.
- In case of real or alleged malfunctioning, neither tamper with the device nor try to repair it by yourself.
Neither intervene on the device nor disassemble it. Only specialized and authorized centers can repair the device.
- Avoid violent impacts that may damage the device.
- Clean the device only by using disinfectant with sodium hypochlorite or quaternary ammonium salt (percentage: 0.2-0.3%)
diluted with distilled water. After cleaning and disinfecting, dry the device and its accessories with a clean cloth.
- It is recommended that you clean/disinfect the parts after every use, unless otherwise indicated.
Always use the device with clean hands.
- It is recommended that you use the device in a clean room to avoid the contamination of the device with dust and dirt.
- It is recommended that you use the device in a ventilated, well-aired space.
- It is recommended that you clean/disinfect after every use.
- It is recommended performing the maintenance of the device and its parts every 12 months since the date of purchase.
- Place the device in its original packaging when it is not used or has to be transported.
Disposal of the device
Do not throw the device or part of it in the fire, but dispose of the product in the specialized centers and respecting the
directives in force in your country. When the product has to be disposed of, the user can give it back to the retailer when
purchasing a new device.
Correct waste separation and returning the end-of-life product contribute to avoiding negative impacts on the environment
and health and promote the reuse and/or recycle of materials. The illegal disposal of the product involves the application of
an administrative fine according to current regulations.
The motor point pen is a very useful tool for search as precisely as possible the best place to position the electrode and get
the most complete contraction.
-
Connect the device to one of the two plugs at the end of the cable for electrode connection of the Globus stimulator.
-
Connect an electrode to the other plug to stimulate (chosen according to the size recommended for the treatment
that you plan to carry out).
-
Place the electrode on the muscle area to be stimulated, as suggested in the section on positioning electrodes in
the operating manual of the stimulator.
-
The area on which to apply the search for the motor point is the area corresponding to where it is recommended
the positioning of the other electrode. It is advisable to prepare the area by covering it with gel.
-
Place the tip of the pen on a point of the area previously identified.
-
Select the motor point pen program from the menu of the stimulator (alternatively, one of the warm-up programs).
-
Raise the intensity of the channel on which the cable is connected. The recommended intensities are:
4-6 mA for the muscles of the upper limbs and trunk;
5-9 mA for the muscles of the lower limbs.
-
Slowly move the pen and remain within the electrode area that will be used later, searching for the point at which
we will feel the maximum increase of the contraction. This will be the best point where to place the electrode.
MAINTENANCE AND CLEANING