Digitize Dc Current - Fluke 8588A Service Manual

Reference multimeter / 8 1/2 digit multimeter
Hide thumbs Also See for 8588A:
Table of Contents

Advertisement

Digitize DC Current

Configuration: Aperture = 200 ns, Trigger count 101010, Low pass filter Off.
Range
Reference Value
10 mA
0.000 000 00mA
10 mA
10.000 000 mA
10 mA
-0.000 000 mA
10 mA
-10.000 000 mA
Configuration: Aperture = 200 ns, Trigger count 101010, Low pass filter 100 kHz.
Range
Reference Value
10 mA
0.000 000 0 mA
10 mA
1.000 000 0 mA
10 mA
-0.000 000 0 mA
10 mA
-1.000 000 0 mA
Configuration: Aperture = 200 ns, Trigger count 101010, Low pass filter 3 MHz.
Range
Reference Value
10 mA
0.000 000 0 mA
10 mA
1.000 000 0 mA
10 mA
0.000 000 0 mA
10 mA
-1.000 000 0 mA
Permissible Error (±)
8588A
0.001 600 mA
0.002 620 mA
0.001 600 mA
0.002 620 mA
Permissible Error (±)
8588A
0.000 900 mA
0.001 520 mA
0.000 900 mA
0.001 520 mA
Permissible Error (±)
8588A
0.000 900 mA
0.001 520 mA
0.000 900 mA
0.001 520 mA
Reference Multimeter / 8½ Digit Multimeter
8558A
0.001 600 mA
0.000 280 mA
0.002 620 mA
0.000 280 mA
0.001 600 mA
0.000 280 mA
0.002 620 mA
0.000 280 mA
8558A
0.000 900 mA
0.000 230 mA
0.001 520 mA
0.000 240 mA
0.000 900 mA
0.000 230 mA
0.001 520 mA
0.000 240 mA
8558A
0.000 900 mA
0.000 230 mA
0.001 520 mA
0.000 240 mA
0.000 900 mA
0.000 230 mA
0.001 520 mA
0.000 240 mA
Calibration Verification
Fluke Calibration
Uncertainty (±)
Fluke Calibration
Uncertainty (±)
Fluke Calibration
Uncertainty (±)
15