DC
neroid Sphygmomanometer
(manometer, cuff, bladder, bulb and valve)
Thank you for choosing an DC
DC
®
DI GNOSTIX
comfort. Every component has been carefully designed to maximize performance. This
Use, Care, & Maintenance guide refers to DC model series; 700, 703, 705, 720, 728, 731,
732, 740, 750, 752, 778, and 788.
Device Description and Intended Use
DC
®
aneroid sphygmomanometers are used by professional healthcare providers and in-
dividuals trained in the auscultatory blood pressure technique to determine systolic and di-
astolic blood pressure in humans.
Contraindications:
neroid sphygmomanometers
are contraindicated for neonate use.
Do not use with neonatal cuffs or
neonate patients. Review the size
chart (right) for proper age and
limb range usage.
General Warnings
warning statement in this manual identifies a condition or practice which,
if not corrected or discontinued immediately could lead to patient injury, ill-
ness, or death.
W RNING: Do not allow a blood pressure cuff to remain on patient for more than 10
minutes when inflated above 10 mmHg. This may cause patient distress, disturb blood
circulation, and contribute to the injury of peripheral nerves.
W RNING: If luer lock connectors are used in the construction of tubing, there is a
possibility that they might be inadvertently connected to intra-vascular fluid systems,
allowing air to be pumped into a blood vessel. Immediately consult a physician if this occurs.
W RNING: Safety and effectiveness with neonate cuff sizes 1 through 5 is not estab-
lished.
W RNING: If this equipment is modified, appropriate inspection and testing must be
conducted to ensure its continued safe use.
®
DI GNOSTIX
TM
aneroid is designed to provide unrivaled durability, and unparalleled
TM
neroid Sphygmomanometer. Your new
Size Chart
Cuff
Size
Infant
7
Child
9
Sm. dult
10
dult
11
Lg. dult
12
Thigh
13
Limb Range
Inches
CM
3.5 to 5.5
9 to 14
5.1 to 7.6
13 to 19.5
7.4 to 10.6
19 to 27
9 to 15.7
23 to 40
13.3 to 19.6
34 to 50
15.7 to 25.9
40 to 66
Need help?
Do you have a question about the Diagnostix 700 Series and is the answer not in the manual?