Pursuant to Medical Device Directive 93/42, the apparatus must only be operated by
trained personnel. Prerequisite is an exact knowledge of the operating instructions!
The F. STEPHAN GmbH will not be liable for damages caused by faulty operation due to
non-observance of the operating instructions or from use for which the apparatus is not
earmarked.
Inspiration
0,25 - 2,00s
Expiration
0,25 - 30,00s
manual ventilation with
prefixed patterns PEEP
PLATEAU
manual inflation
fill bottle with Aqua Dest.
(sterile and demineralised
water) not over max. level!
Reduce level, if flow is
higher than 10l/min!
Water is sufficient for 4-6 h
Oxygen flowmeter for aux
outlet for incubator
aux. O2 connector for
incubator
O2 sensor in normal posi-
tion removable for use in
incubator. Close the socket
with replacement plug if cell
is outside (prevents high
leakage flow)
brief instruction F120 GT ver 1.0
23.08.02 JS
Reanimator F 120 Infant Transport System
Display of frequency
three „---" in CPAP mode
blinks if out of possible
range max. 120/min (IMV
Mode)
9
2
2 min alarm
cable alarm
suppression
blinks at failure of patient
for NOVA
cable or fixation of elec-
trodes
Turn OFF:
Press
key,
press additional
key,
unhand red
key
GLOBETROTTER
trained personal only
IMV
default mode at start
8
3
0
5
0
0
0
5
0
15
100
200
kPa x 100
kPa x 100
0
300
bar
bar
315
STEPHAN
STEPHAN
®
®
menu key to
move up or
down
patient alarm
blinks at cardiac or
respiratory alarm
®
2 min
suppression
LED for
Stenosis Visual alarm if the
permanently installed pressure
of 12 mbar is not fallen below 2,2
seconds after the set end of inspi-
ration time (stenosis)
Disconnection Visual alarm
ensues if the permanently installed
pressure limit of 12 mbar is not
execeeded after the set expiration
time +max inspiration time (2.0s)
20 30
O2 Alarm Visual alarm in case of
10
40
STEPHAN
®
violation of the set limit values for
0
50
cmH O
-10
60
O2 concentration (monitor must be
2
Überlastgrenze: 600 cmH O
2
switched ON)
Beatmungsdruck
Battery Visual and acoustic alarm
if internal battery voltage falls
L/min
1.013 bar
10
10
20°C
below 10,8V. Heater is off.
Every 30s warning tone occurs not
suppressable. 20 minutes remai-
5
5
ning then respirator is shut off
1
1
Med.
AIR
O2 pressure
cylinder/
content
O2 pressure
internal
Battery only
NOVA
mains alarm
only NOVA