Health O Meter PRO PLUS 1100KL Operation Manual page 34

Electronic scales
Table of Contents

Advertisement

Available languages

Available languages

ESC
character
Name
Response
P
Diagnostics(1)
A
Diagnostics(2)
Z
Control (set a
C
value)
Get subject
G
record data and
readings
Set subject
S
ecord data
Clear readings
X
Clear subject
Y
record data
Name of Control
Auto Hold Time
Auto Off Time
Sleep Time
Keypad Volume
Stable weight Volume
Alarm Volume
Display of Date/Time
Backlight
Contrast
Weight Display Mode (fractions(f), decimals(d)
or Pounds and ounces (p)
Tare Weight
Unit Of Measure (Metric or constitutional)
Positions left in RAM
Table A - Scale Global Values List and Identifiers
Name of component to test
Load Cell A
Load Cell B
Load Cell C
Load Cell D
Battery
PC Communication (USB)
Write to RAM
Read from RAM
Table B - Components to test
ESC value
with
parameters
This will be the scales response to the request mentioned above
Pccc=ccc
cccc
Accc
This is the request for a diagnostic test on certain parts of the scale
(such as battery life, loadcells, keypad, LCD etc.). (See Table B)
This will be the response of the diagnostics done on the scale.
Zccc
Values will include error codes to indicate what is wrong with the
scale, or all zeros to indicate that all is well.
Cccc=ccc
his is to set the value of the scale's global settings such as backlight,
ccc
auto hold time, keypad beep volume etc. (See Table A) (i.e.
<ESC>CAHT=20<ESC>E will set the Auto Hold Time to 20 seconds)
This will return all the information from both the subject record data
G
as well as the readings stored in the readings table.
This will set the subject record data for a specific patient.
S
Immediately following this will be the value that is set. (i.e.
<ESC>S<ESC>I1234<ESC>T200<ESC>Um<ESC>E will set the
TARE for patient ID1234 to 200kg)
This will clear all the values stored in the readings table on the scale.
X
Only the subject record data will not be deleted.
Y
This will clear the subject record data for a specific patient. The
patient ID will follow the Y immediately (<ESC>Iccccccccccc). If no
ID is given, ALL the data will be cleared.
Identifier
AHT
AOT
STM
VKE
VSW
VAL
DAT
BAC
CON
WDM
TAR
UOM
RAM
Identifier
Error Code
LCA
E3A
LCB
E3B
LCC
E3C
LCD
E3D
BAT
E4L(Bat Low, but still usable) or
E4U (Bat Low and Unstable)
PCC
E05
WRM
E08
RRM
E09
Description
Unit
nn (Seconds)
nn (Minutes)
nn (Seconds)
nn (01..10)
nn (01..10)
nn (01..10)
nn (01 or 10)
nn (01..10)
nn (01..10)
c (f,d,p)
nnn.nn (Kg)
c (m or c)
nnnn (0..max)
33