Page 1
No light is illuminating the sample..........3 The Control Pod display appears blank, but functionality remains. 4 The light guide doesn’t fully insert into the Light Source (light guide version only)................ 4 CoolLED contact information ............4 DOC-037 Iss 1...
Page 2
1) The Illumination System doesn’t power up (nothing appears on the Control Pod). fura Is the power connector fully inserted into the pE-340 Light Source? Is the mains lead fully inserted into the power supply? Is the mains supply switched on? ...
Page 3
If the offset moves with the rotation of the Light Source then the system has lost its alignment. This will need to be returned to CoolLED to be reset. If the offset does not move with the rotation of the system then the issue is with the optical path of the microscope.
Page 4
If the advice in this document has been unable to solve your issue, or the issue that you are experiencing is not listed then please contact the CoolLED support team at support@coolled.com. They will be able to offer further advice and explain the repair procedure.