Product Manual The Safety Team's Resource for Setup, Installation, Operation, and Service Edition: 5 November 1, 2019 Part number: 17158071-1...
Contents Warnings and Cautionary Statements ......................v Certification Summary ..........................vii European Union Radio Equipment Directive ................... viii Chapter 1 ............................... 1 Product Information ............................1 Product Overview ............................1 System Overview ............................1 Key Features ............................. 3 RGX Gateway ............................3 Communication ............................
Page 4
Supplies .............................. 39 Instructions............................39 Warranty ..............................46 Limitation of Liability ..........................46 Appendix A ..............................47 Requirements for electrical connections (Industrial Scientific Control Drawing 1810D9509-200) ... 47 Appendix B ..............................51 Applicable Certification Standards ......................51 Contact Information ............................. 52...
Page 5
Tables and Figures Table 0.1 RGX Gateway hazardous location certifications ................vii Table 0.2 RGX Gateway wireless certifications .................... vii Table 0.3 EU RED radio transmitter types ....................viii Figure 1.1.A System overview ........................2 Figure 1.1.B System overview with RGX in LENS repeater mode ..............2 Table 1.1 Feature Comparison –...
Page 6
Figure A.1. B External electrical connections in a nonhazardous location ........... 48 Figure A.1.C Internal electrical connections ....................49...
WARNING: Connect or disconnect only in a non-hazardous area. CAUTION: Battery pack is only Industrial Scientific technician replaceable with ISC P/N 17157552; use of another battery may present a risk of fire or explosion. The battery used in this device may present a risk of fire or chemical burn if mistreated: Do not crush, disassemble, or incinerate.
Page 8
Industrial Scientific recommends persons with a pacemaker or implantable cardio defibrillator (ICD) should maintain a minimum separation distance of 15 cm (6 ") between the pacemaker or ICD and a wireless-enabled device.
See Appendix B for a List of Applicable Certification Standards. In addition to the certified wireless uses summarized below, refer to the Industrial Scientific website for the most up-to-date information on wireless product certifications. Table 0.2 RGX Gateway wireless certifications...
Page 10
EU Radio Equipment Directive The RGX Gateway has been assessed and meets the requirements for the Radio Equipment Directive (CE RED). To determine if your unit is CE compliant refer to the unit’s label. The RGX can utilize the following radio transmitter types: Table 0.3 CE RED radio transmitter types GPS Receiver Module...
*Linked Equipment Network for Safety. System Overview LENS Wireless is a long-range, power-efficient wireless mesh network from Industrial Scientific. The exchange of LENS group data supports the live-monitoring capabilities of iNet Now. From a computer or smart-device, iNet Now users can learn, on a live basis, about everything from instrument gas readings to gas alarms, man-down events, panic alarms, and more.
iNet Instrument data RGX Gateway iNet Now users (live monitoring and real-time alerts) Figure 1.1.A System overview iNet Main facility iNet Now users RGX Gateway (live monitoring & real-time alerts) RGX Gateway in LENS repeater mode extending the facility’s LENS Wireless group to another building.
• More than one RGX can be included in a LENS group. • RGX units operating in LENS repeater mode also use this network. Instrument data exchanged through LENS are automatically encrypted with the Industrial Scientific key. Optionally, a customer-supplied encryption key may be used.
Page 14
RGX Gateway • Gateway critical error • Gateway online • Gateway offline Noncritical events: Set the interval value for noncritical events in iNet from 1 to 5 minutes. Gas detection instruments • Pump alarms • Peer instrument system alarms • Peer instrument calibration events •...
• The RGX can receive operational power from a compatible power-supply accessory only from Industrial Scientific; use of each accessory requires the RGX to be equipped with its Intrinsic Safety (IS) Cable Adapter. This option is well suited for use conditions that do not require permanent installation but demand a run time exceeding that of the factory-installed battery.
The operational run time for the factory-installed battery can be extended with the use of a compatible power-supply accessory only from Industrial Scientific. Each power-supply accessory, as listed below, has unique use restrictions or certified uses. Before using either accessory, read and understand its product...
Table 1.4 RGX Gateway intrinsic safety (IS) input parameters Item Value 16.2 VDC 0.980 A 2.2 W 0 µF 0 mH The RGX Gateway can be operated using customer-supplied, customer-installed fixed DC power that supplies a compatible input voltage range of 9–30 VDC with a maximum current of 5.0A connected to the dedicated ports on the RGX terminal block.
Table 1.5 RGX Gateway specifications Item Description Power input Three mutually exclusive options: • Terminal block: 9–30 VDC, 5.0A (Hazardous Locations) • 12V power supply: 12VDC, 5.0A (Nonhazardous Locations Only) • IS power: 16.2 VDC, 0.980A Size (length x width x depth) 28 cm x 23 cm x 15 cm (11 ″...
The RGX Gateway can draw operational power from compatible power-supply accessories only from Industrial Scientific. The RGX can run indefinitely on either the Intrinsically Safe Extended Run Time Power Supply (ISERTPS), Industrial Scientific Product Manual part number 17158248, or the Extended Run Time Power Supply (ERTPS), Industrial Scientific Product Manual part number 17158385.
Case lid Case base Fastener (4x); Handle 1.69 N m; 240 oz in screw torque Hinge Screw hole (4x); for use with optional Wall Mount kit or Power button Magnet Mount kit Indicator light (system-power- communication status) Indicator light quick reference Labels NFC target...
Chapter 2 Deployment Planning Introduction Communications Settings and Connections Introduction Whether you are using a single RGX™ Gateway or multiple units, this chapter will help you make decisions about how each will operate within your organization. An RGX Gateway operating in LENS™ repeater mode connects LENS group peers (instruments and RGX Gateways) maintaining wireless connectivity among peers without any additional networking equipment.
"X" is used here to indicate the value will be a single character in length; the available values are the letters A through J. "Default" sets the unit's LENS Wireless to the Industrial Scientific encryption key, and "custom" to the customer's key. If a custom key will...
You will need to make decisions about the RGX GPS and noncritical data settings. • Determine if the unit will operate with its GPS on or off. When on, the unit can acquire its GPS coordinates and send this location data to iNet. Importantly, the GPS data are used by iNet Now users to identify the location of an RGX that has sent data from an in-alarm gas detector.
During the unpacking process, account for and examine each ordered item. If any item is missing or appears to have been damaged, contact Industrial Scientific or an authorized distributor of Industrial Scientific products. After unpacking, peel the protective plastic film from the case lid.
The values noted below, unique to each RGX, are required for setup. Side (right) Case base Serial number (S/N) Required for unit activation. Activation code Required for unit activation. MAC addresses (on label) Required for Ethernet configuration and permanent installations. IP Address Note: If the value is unknown to you, see your network administrator who may need the unit's MAC...
Charging the battery Using only the supplied RGX charging power supply and power cord, charge the unit's installed battery as shown and described below. Nonhazardous location Charging power supply port Indicator light (charging status) Charging power supply and power cord ─...
Power On and Shutdown Locate the power button on the case lid. Power on Press the power button for approximately 5 seconds until the power-button light turns on. • If the unit is fully operational, the power-button light will be solid (not blinking) green. •...
There are two methods for accessing the RGX Configuration Interface—through a wi-fi or Ethernet connection. Supplies required for each access method are listed below. Table 3.3 Supplies needed for configuration Access method Item Wi-fi Ethernet Torx screwdriver with a T25 bit Ethernet cable Cat5 or greater Computer (with browser) Smart device (with browser)
Page 32
Figure 3.3 Power on the RGX On the case lid, press the power button for approximately 5 seconds until the power-button light turns on. Observe the power-button light and proceed as noted below. Green only (blinking or solid) Continue. Includes red (blinking or solid) Table 3.6 Power-button light indicators before continuing.
RGX Gateway operating in LENS repeater mode Site name can be set. All other settings must be edited in iNet. Save and Reboot When finished, click save; the Interface will prompt you to confirm the save. When prompted to reboot, click on the prompt, then confirm. No other action is needed, the software will complete the reboot.
Page 34
Connect Ethernet cable; power on unit Before connecting the Ethernet cable, you will need to ensure the computer and the RGX are on the same network. Consult your network administrator for assistance. Note: For network connections, the default static IP is 192.168.1.1. Next, connect one end of the customer-supplied Ethernet cable (Cat5 or greater) to the computer, then connect the other end to the Ethernet port.
Ethernet cable from the unit. • If there are errors or issues, recheck the work completed to this point. If an issue persists, contact Industrial Scientific. • Disconnect the cable from the computer and, as needed, reconfigure the computer to the desired network.
Table 3.5 Internal indicator lights* and their meanings Label Indicator Light Operational status COMM STAT iNet connection status Green The unit has a connection to iNet; if blinking green, data are being exchanged. The unit has no connection to iNet. LENS STAT LENS Wireless status Green...
Log into iNet Follow this navigation. • Click on the Equipment menu. • Choose Equipment List. At the equipment list page (not shown), use the search box associated with the Serial Number column to enter the unit's serial number. Click on the unit's serial number; the unit's information will display. Configure LENS Wireless and General settings Go to settings.
See also for more information. An error has occurred that requires resolution. Power off, then power on the unit. If the issue persists, contact Industrial Scientific The unit is powered off or the battery is completely discharged. Light is off...
Chapter 4 Installation and Operation Introduction Site Selection Placement and Mounting Installation Preparation Installation Operation Precheck Introduction Only qualified personnel should install and operate the RGX™ Gateway. To help prevent injury and damage to the equipment, handle the unit with care and avoid dropping it. NOTICE: A customer's intended use of the RGX may require mounting, as well as the installation of a compatible fixed DC power source, an Ethernet connection, or both.
Magnet Mount Kit from For use with metal surface Do not use with units that will draw power from a fixed DC Industrial Scientific mounting power source. See the Service section of this manual for instructions.
AHJ. Ensure a minimum distance of 5 m (16 ' 4 ″) between a unit and its optional power-supply accessory from Industrial Scientific, or any that are in use nearby.
ISERTPS RGX Gateway 5 m (16 ' 4 ″) (3 ' 3″) Ground Figure 4.1 Distance requirements: RGX-to-power-supply and RGX-to-ground ISERTPS shown, also applies to ERTPS. Also applies to an RGX Gateway operating in LENS repeater mode. Placement and Mounting The unit can be placed or mounted for operation.
Table 4.2 Possible installation preparation tasks Application Service task Conversion from fixed DC power source to any other Acquire the Conduit-hub Plug Kit and complete installation. Do not operational power source reinstall any used conduit-hub plug components; once removed a plug's seal is compromised and other plug components may become altered or damaged.
If your installation excludes connecting both fixed DC power and Ethernet, Figure 4.2 continue below with step 1; otherwise, use the instruction set in Figure 4.3. ─ Mount or place RGX If the unit has been prepared for mounting, use the customer-supplied equipment and services to complete the mounting task;...
Page 45
If the RGX will be connected to an external power source*, de-energize the external power source. *A customer-supplied and customer-installed fixed DC power or a compatible power-supply accessory from Industrial Scientific. Shut down the RGX Observe the RGX power-button light.
Page 46
Access case interior When working inside the unit, avoid contact with the battery, other exposed components, and any item protected by the unit's interior plastic cover. Avoid dropping or spilling anything into the case. • Using a torx screwdriver with a T25 bit, unscrew and remove the four fasteners that secure the case lid to the case base.
• Use these mechanical and electrical wiring specifications. Mechanical: Solid Wire: 18-12 AWG Stranded Wire: 18-12 AWG Torque: 4.0 lb in (0.45 N m) Electrical: Shield wiring, or conduit, at the source Connect optional Ethernet cable If an Ethernet connection will be used, connect the cable to its port. Secure the cable and power line Use the guides to secure the already connected Ethernet cable and power line.
Use protective gloves and eyewear to avoid a potentially severe pinch injury, cut, or splinter. Service tasks that can be completed by Industrial Scientific customers are described in this manual. Table 5.1 indicates which parts and components are customer replaceable. All other service tasks should be performed only by Industrial Scientific or its authorized service centers.
Figure 5.1 Replaceable parts diagram for the RGX Gateway Table 5.1 RGX Gateway customer replaceable parts list Diagram Part name Part Notes number number Case fastener 18109573 Four fasteners per pack Fastener torque: 1.69 Nm (240 oz in) Charging port cap 18109570 Includes tether-chain, screw, sealing gasket, washers and hex nut.
Page 51
Figure 5.2 Instructions for service items installed on the case exterior: mounting kits and IS cable adapter. Wall Mount Kit installation — Using the orientation shown, mate the hexagon cutout of the wall Using a T30 torx screwdriver, Install the remaining wall mount to a corresponding corner on the back of the RGX.
Intrinsic Safety (IS) Cable Adapter installation IS Cable Adapter RGX IS port connector Arrow Wide notch Turn the IS power port cap Align the arrow on the top of the adapter connector to the wide notch on the IS port; push on to counterclockwise the port to connect.
Page 53
Figure 5.3 Instructions for service items that require interior case access: conduit-hub plugs, vent, and port caps. Opening the case Closing the case Use a T25 torx screwdriver to The unit is hinged on the left; When closing the lid, ensure all wires and cables are secure unscrew and remove the four open the lid from right to left.
Page 54
Conduit-hub Plug Kit installation Use only an unused Conduit- Inside case hub Plug Kit from Industrial Hex nut with nylon insert Scientific. Flat washer Do not reinstall any used conduit-hub plug components; Sealing gasket once removed, a plug's seal is Washer compromised and other plug Fixed gasket...
Vent replacement To remove the vent, use an 18 mm open-end wrench to hold Insert the replacement vent into the hole from outside of the the vent in place. case. Inside the case, turn the nut counterclockwise with a 3/4 ″ Inside the case, screw the nut onto the vent threads clockwise.
This warranty includes batteries, except where otherwise stated in writing in the literature accompanying the product. Industrial Scientific warrants that it will provide a cellular data plan for the life of the Gateway, for a minimum of two years from time of purchase provided the customer maintains an active iNet® Now account during that time;...
Appendix A Requirements for electrical connections (Industrial Scientific Control Drawing 1810D9509-200) From Figure A.1.B Figure A.1.A External electrical connections in a hazardous location...
Page 58
Figure A.1.A Figure A.1. B External electrical connections in a nonhazardous location...
Appendix B Applicable Certification Standards • UL 121201; Nonincendive Electrical Equipment for Use in Class I and II, Division 2 and Class III, Divisions 1 and 2 Hazardous (Classified) Locations; 9 • UL 61010-1; Electrical Equipment for Measurement, Control, and Laboratory Use; Part 1: General Requirements - 2 Ed.;...
Vertriebshändler unserer Produkte oder ein Servicecenter bzw. eine Niederlassung von Industrial Scientific zu finden. Para buscar un distribuidor local de nuestros productos o un centro de servicio u oficina comercial de Industrial Scientific, visite www.indsci.com. 如需查找就近的产品经销商或 Industrial Scientific 服务中 心或业务办事处,请访问我们的网站 www.indsci.com...