  • Page 2 Thank you for purchasing the Signature Series Model OC-2300 Ophthalmic Chair. Please read this User's Manual carefully and place it in a convenient location for future reference. WARNING: The installation of the chair shall be performed only by a qualified technician.
  • Page 3: Table Of Contents

    CONTENTS 1.0 NOMENCLATURE 1.1 Model OC-2300 Main Components 2.0 PERFORMANCE AND SPECIFICATIONS 3.0 ASSEMBLY Tools Ophthalmic Chair / Stand Position Electrical Connections Alternate Connections Power Up 4.0 OPERATION Elevation Control Motorized Recline Control Auto-Return Control Memory Controls Cancel Button Swivel Brake Operation Armrest Retraction Headrest Position Footrest Retraction...
  • Page 4: Nomenclature

    NOMENCLATURE Signature Series Model OC-2300 Main Components Figure 1 shows the primary features and key components of the OC-2300 Ophthalmic Chair. Use the figure as a reference. Headrest Chair Controls (both sides) Armrest Recline Position Brake Handle (both sides) Footrest PERFORMANCE &...
  • Page 5 Environmental Operating Conditions • Ambient Temperature Range: 50 to 104 Degrees F (10 to 40 Degrees C) • Relative Humidity Range: 30 to 75% • Atmospheric Pressure Range: 10.1 to 15.4 psi (700 to 1060 hPa) Environmental Storage/Transportation Conditions • Ambient Temperature Range: -20 to 130 Degrees F (-23 to 54 Degrees C) •...
  • Page 6: Assembly

    Safety Switch Blue WARNING: If the chair is used with a Topcon IS-800 stand, the plastic base shroud (non-fuse side) must be removed to make a configuration change on the circuit board. This will make the IS-800 safety switch feature operable.
  • Page 7: Power Up

    If the chair has a 1 generation circuit board, remove the blue jumper pin located between the transformer and the 9-pin connector. Re-install the base shroud. If the chair has a 2 generation circuit board, move the red jumper pin located between the 9-pin connector and the 6-pin connector labeled ‘Stand Safety’...
  • Page 8: Memory Controls

    Memory Controls 4.4.1 Programming Press the AUTORETURN button to position the chair in the fully down and fully upright position (home position). Using one of the membrane switches located in the sides of the chair backrest, move the chair to a desired position. Press and hold one of the memory buttons for 3 seconds.
  • Page 9: Adjustments

    ADJUSTMENTS Headrest Locking Tension If the position of the chair headrest has begun to move with the clamp-locking handle in the locked position, the headrest clamp may require an adjustment. Follow the instructions below to adjust the headrest assembly. Headrest Clamp Adjustment 5.1.1 Release the clamp-locking handle.
  • Page 10: Maintenance

    set screws in the appropriate armrest stop to align the armrests with one another. Reinstall the backrest upholstery after performing the adjustment. MAINTENANCE It is recommended that the exterior surfaces of the OC-2300 chair be cleaned periodically to preserve the chair's appearance. Cleaning should be performed at least every six months.