ChemInstruments AR-2000 Operating Instructions Manual page 19

Adhesion/release tester.
Table of Contents

Advertisement

MODE
The AR-2000 can be operated in a constant speed mode or a ramp mode. The
following is a step-by-step procedure for setting the mode.
1. Select SETUP from the main screen.
2. Select MODE from the setup screen.
3. Select the desired mode. Select OK to confirm the entered mode.
SETTING SPEED
NOTE: The unit of measure for speed can be changed by selecting UNITS from
the setup screen.
Constant Speed Mode
To perform a test correctly, it is necessary to set the sled speed in accordance with the
selected test method. The following is a step-by-step procedure for setting the sled
speed in constant speed mode.
1. Select SETUP from the main screen.
2. Select SPEED from the setup screen.
3. Select CHANGE and enter the desired sled speed in the selected units and press
ENTER. Select OK to confirm the entered speed.
Ramp Speed Mode
The following is a step-by-step procedure for setting the ramp speed mode parameters.
1. Select SETUP from the main screen.
2. Select RAMP SETUP from the setup screen.
3. Select the box next to each parameter to enter a screen which allows you to
change the value of the parameter. Select ENTER or OK to confirm the
parameter.
Adhesion/Release Tester Operating Instructions (AR-2000)