ADA5000
Automatic Distillation Analyzer
Operation and Instruction Manual
REV A
Koehler Instrument Company, Inc.
1595 Sycamore Avenue • Bohemia, New York 11716-1796 • USA
Toll Free: 1-800-878-9070 (US only) • Tel: +1 631 589 3800 • Fax: +1 631 589 3815
http://www.koehlerinstrument.com • e-mail: info@koehlerinstrument.com
Petroleum Testing & Analysis Instrumentation • Custom Design & Manufacturing

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ADA5000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Koehler ADA5000

  • Page 1 Automatic Distillation Analyzer Operation and Instruction Manual REV A Koehler Instrument Company, Inc. 1595 Sycamore Avenue • Bohemia, New York 11716-1796 • USA Toll Free: 1-800-878-9070 (US only) • Tel: +1 631 589 3800 • Fax: +1 631 589 3815 http://www.koehlerinstrument.com •...
  • Page 3: Table Of Contents

    ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual Table of Contents 1 Int roduction ..................1 Koehler’s Commitment to Our Customers ........... 1 Recommended Resources and Publications........1 Instrument Specifications ..............2 Software Specifications ............... 2 2 Safety Information and Warnings ............2 3 Unpacking &...
  • Page 4: Introduction

    IP Publication: requirements and provide you with the best  IP 123: solutions. You can depend on Koehler for a full range of accurate and reliable instrumentation 4. Deutsche International Norm (DIN) as well as support for your laboratory testing http://www.din.de...
  • Page 5: Instrument Specifications

    ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual specified or implied, and may result in personal injury and/or property loss. Replacement parts 1.3 Instrument Specifications must be O.E.M. exact replacement equipment. Models: K45603, K45604 K45703-TS, K45704-TS Internal Access. If internal access into the instrument is necessary, the analyzer must be Electrical Requirements: 120VAC ±...
  • Page 6: Unpacking & Installation

    If this does not happen within 10 shipping container and equipment. Do not return seconds, then browse the files on the CD-ROM goods to Koehler without written authorization. and double click on the setup file (setup.exe) to start the installation. Follow the instructions on Instrument Installation the screen to setup the software.
  • Page 7 CD. Continue by clicking on the “Next” button. Click on the “Next” button. Utilize default directory C:\Program Files\Koehler. Click on the “Next” button to continue. ADA5000-Manual - 4 -...
  • Page 8 ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual Utilize the default data directory C:\Program Review the program settings. Click “Next” to Files\Koehler\data. Click on the “Next” button to continue. continue. Then, select “Custom” to copy all programs and a demo database. Click the “Next”...
  • Page 9 Operation and Instruction Manual Select the “General” tab at the top of the screen. It is not recommended to alter the various paths; please consult with Koehler technical service. To change the units for temperature pressure (2), select the desired units by clicking on the corresponding radio buttons.
  • Page 10: Operation

    2. Software. After the PC has been properly booted, start distillation software The Koehler Automatic Distillation Analyzer is program by double clicking on the Koehler designed for acquiring test data in accordance icon located on the computer desktop and with ASTM related test set up the parameters as needed.
  • Page 11: Communication Manager

    5. To run the software program, double 20. Select “Save“ to store the first run and click on the Koehler icon on the select “Print“ to print out the results to computer desktop. the printer.
  • Page 12 ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual To observe the data transfer process from the PC to the analyzer, select the “View” option and then “Data Transfer”.  Select unit (1). Select a connected analyzer to see all the important values of this analyzer.
  • Page 13: Distillation Control

    ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual Distillation Control This program window will appear after starting “distctnl.exe”. It is possible to change the path of the database, program form and results form. For a normal Create new program installation you can use this standard directory.
  • Page 14 ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual “Dist.Data“. This point has no influence if using Results Comparison the printer. To compare up to 4 distillation results, select the Cooling Dev.: graph button. You can compare temperature, Compare cooling device for more detailed heat, and time for up to four sets of data.
  • Page 15: Current Program

    ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual Current Program Save Select item to store program into the database. Unit Select analyzer Load Select item to load a new program from the database. Attention: Program will be sent automatically to selected analyzer and overwrite an existing previous run.
  • Page 16 ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual Print Select item to print out program of selected analyzer. To change form, go to Config then ProgramForm. Sample Print Out Program Diesel Sample Diesel Sample ID 12.9.00 Operator Koehler Comment Sample...
  • Page 17 ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual Method Opt Heat #2 Select method (ASTM D86, D1078, D850, DIN If distillation run has finished, (min. 30 mL are 51751) necessary) microprocessor system of distillation analyzer will calculate optimized value to get desired distillation rate for the first 2 mL.
  • Page 18 ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual Stop at Dry Point Select “No“ to stop distillation run after detecting dry point, or “Yes“ to wait for FBP. Solvents off Time Look at info on dry point on how to set this point.
  • Page 19: Diskstat.exe

    Temperature: Temperatures in °C or °F are in any chosen directory. In a normal case use possible. Please note, that program works the local directory. Koehler software can also be internally with °C and all temperatures stored in used in a network system.
  • Page 20: Commgr

    Grid Steps – PrintForm – PrintFormHeader located under or near ceiling mounted sprinkler type fire extinguishing heads. The area above Data Base: The Koehler software uses a the distillation unit should be fireproof also. From standard ACCESS database to store programs time to time, especially after a fire, the sensor and results.
  • Page 21: Maintenance

    If you have any questions about maintaining your equipment, please do not hesitate to contact the Koehler technical service department. Routine Maintenance The Koehler Automatic Distillation Analyzer...
  • Page 22: Service

    Warranty number prominently displayed will be returned to the sender. Goods must be returned freight We, at Koehler, would like to thank you for prepaid. Returns will be subject to a restocking your equipment purchase, which is protected by charge, the application of which will depend the following warranty.
  • Page 23 ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual Notes ADA5000-Manual - 20 -...