To all operators and users of
Babytherm 8004/8010 open care units
for premature and newborn patients
lmportant safety notice!l!
Babytherm 8004/8010
open care units for premature and newborn patients
Dear Madam/Sir,
Dräaer
June
2016
As part of our market and product monitaring we have become aware of a case in which
a Babytherm 8004 without patient was operated in skin temperature mode with skin
temperature sensor to prepare it for use. ln skin temperature mode, the radiant warmer
output
is
controlled automatically until the set skin temperature
is
reached. As a result of
operating the temperature sensor without patient contact,
the
radiant warmer was
operated continuously at a high output Ievel. During this forced continuous operation
under high thermal Ioad, the heating element aged faster. Apart became detached from
the heating element and fell onto the mattress tray
\1\fhere
a towel caught fire.
Do not operate the device in skin temperature mode with temperature sensor
without patient contactl
To preheat the device, you can use the manual mode up to Ievel 3 as there
is
no active
time Iimit. At Ievel 4 or above, there will be regular alarms. lf the
alarm
is not
acknowledged, the radiantwarmer will be automatically switched off. For further details,
please refer to the Babytherm 8004/8010
instructions
for
use.
Additionally, our analyses have shown that the heating elements are not replaced at
regular intervals as recommended in the Babytherm instructions for use. We attribute this
to
the fact that it might not be practicable to replace the heating element after 10,000
operating hours. We have therefore converted this value
into
a period of 2 years, which
can be measured more easily.
lf your heating elements have been in operation for more than 2 years, please
replace them.
Drägerwerk
AG
&
Co.
KGaA
Moislinger Allee 53-55
23558
Lübeck,
Germany
Postal
address:
23542
Lübeck,
Germany
TeL +49 451
882.{)
Fax +49 451 882-2080
info@draeger.com
www.draeger.com
VAT ID
DE135082211
Bank
information
:
Company
headquarters: Lübeck
Commerzbank
AG,
Lübeck
Commercial
register:
IBAN: DE95
2304 0022
0014 6795
00
Municipal
Court lübeck
HRB 7903 HL
Swift
Code:
COBA
OE
FF 230
General
partner:
Drägerwerk Verwaltungs
Sparkasse
zu
Lübeck
AG
IBAN:
DE15
2305
0101 0001 0711
17
Company headquarters: Lübeck
Swift
code
:
NOLADE21SPL
Commercial
register:
Municipal Court
Lübeck HRB
7395
HL
Chairman of the Executive Board
for
Drägerwerk AG
&
Co.
KGaA and
Drägerwerk Verwaltungs
AG:
Prof.
Dr.
Nikolaus Schweickart
Executive
Board:
Siefan Dräger
(Chairman)
Rainer Klug
Gert-Hartwig
Lescow
Dr.
Reiner
Piske
Anton
Schrofner

Advertisement

loading

Summary of Contents for Dräger Babytherm 8004

  • Page 3: Important Safety Notice

    To enable us to monitor the effectiveness of this communication, we request that you sign and return (email or post) the 'Confirmation of Receipt' as below. Important Safety Notice Failure of radiant heaters due to overload and ageing Please list affected devices below Part Number Device...
  • Page 4 Babytherm 8004/8010 Ergänzung, Seite 2 Kiegészítő dokumentum, 32. oldal Supplement, page 4 Supliment, pagina 34 Supplément, page 6 Dodatak, stranica 36 Suplemento, página 8 Допълнение, стр. 38 Foglio informativo, pagina 10 Ek, Sayfa 40 pt_ib Documento suplementar, página 12 增补 , 第 42 页...
  • Page 5: For Your Safety And That Of Your Patients

    English English Supplement Babytherm 8004/8010 en Supplement to the instructions for use Babytherm 8004/8010 with part number: 9029458, Keep this supplement with the instructions for edition: 12 – 2015-01 and earlier use of the medical device. Babytherm 8004/8010 (NAWI scale option) with This supplement updates the information of the part number: 9052336, edition: 3 –...
  • Page 6: Technical Data

    5000 oper- ating hours Technical data Radiant warmer Radiant power between the bed and the radiant warmer Heater level 3 max. 10 mW/cm Heater level 10 max. 30 mW/cm 1) measured according to IEC 60601-2-21 Supplement Babytherm 8004/8010...

This manual is also suitable for:

Babytherm 8010