  • Page 1 PTC8001 Temperature Calibrator...
  • Page 2: Table Of Contents

    Table of Contents 1. Introduction ....1 1.1 Customer Service ....1 1.2 Standard Equipment .
  • Page 3: Introduction

    Corporate Office: www.martelcorp.com e-mail: sales@martelcorp.com Tel: (603) 434-8179 800-821-0023 Fax: (603) 434-1653 Martel Electronics 3 Corporate Park Drive Derry, NH 03038 1.2 Standard Equipment Check to see if your calibrator is complete. It should include: Calibrator, Instruction Manual, Test Leads,...
  • Page 4: Safety Information

    1.3 Safety information Symbols Used The following table lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual. Symbol Description AC (Alternating Current) AC-DC Battery CE Complies with European Union Directives Double Insulated Electric Shock...
  • Page 5 Canadian Standards Association The following definitions apply to the terms “Warning” and “Caution”. • “Warning” identifies conditions and actions that may pose hazards to the user. • “Caution” identifies conditions and actions that may damage the instrument being used. Use the calibrator only as specified in this manual, otherwise injury and damage to the calibrator may occur.
  • Page 6 • Remove test leads from the calibrator before you open the battery door. • Inspect the test leads for damaged insulation or exposed metal. Check test leads continuity. Replace damaged test leads before you use the calibrator. • Do not use the calibrator if it operates abnormally.
  • Page 7: Quick Start Instructions

    2. Quick Start Instructions A. Key Functions Function Input/Output Toggles the function selected from measurement mode to source mode. Pressing the “TC” or “Thermocouple” key toggles through all available TC types as well as direct mV output. Pressing the RTD key toggles through all available RTD types as well as direct ohms output.
  • Page 8: Setpoints

    B. Setpoints 1. Storing Setpoints The setpoint mode allows up to 9 setpoints to be set in non-volitile memory for each range. The procedure is as follows: a. Set the output to the desired value. b. Press shift followed by the “SET” key c.
  • Page 9: Connection Diagrams & Instructions

    c. The next screen prompts you to enter the dwell time from 5 to 500 seconds. Simply enter a number in that range. d. Auto-step will begin. e. To exit auto-step press the “CE” key or enter a value. 4. Manual Stepping The PTC8001 can be manually stepped through all or some of the stored setpoints.
  • Page 11: Sourcing A Thermocouple Signal

    Figure 1. B. Sourcing a Thermocouple Signal TC wire must be used to achieve an accurate calibration. Figure 2. Note: For best accuracy allow a 10 minute warm-up period after the PTC8001 is turned on. C. Measuring Temperature Using a Thermocouple Figure 3.
  • Page 12: Measuring Resistance (Rtds)

    D. Measuring Resistance (RTDs) When measuring resistance (ohms) there are 3 choices when using the PTC8001. The choices are 2-wire, 3-wire, and 4-wire measurements. While 4-wire offers the highest accuracy, most industrial RTDs are a balance 3-wire. Figure 4 shows how to hook up the PTC8001 for the 3 types of measurements.
  • Page 13: Sourcing Resistance (Rtds)

    Measuring a 4 Wire RTD Figure 4. E. Sourcing Resistance Into a RTD Transmitter 1. As described previously in section A the PTC8001 can source resistance in a RTD Transmitter or measuring device. In many cases the unit under test will require a 3 or 4 wire connection to achieve best accuracy.
  • Page 14: Maintenance

    4. Maintenance A. Power Requirements The PTC8001 operates on 4 AA alkaline batteries or optional rechargeable Ni-Cd batteries. To replace the alkaline batteries or to install rechargeable batteries remove the two (2) screws on the rear battery door and lift the battery cover. When using Ni-Cd batteries the AC adapter will simultaneously power the PTC8001 and charge the batteries.
  • Page 15: Calibration

    4. If a blown fuse is found replace with the enclosed spare fuse. 5. To order more fuses contact Martel Electronics and order part number 3535039. C. Calibration The PTC8001 should hold its rated specifications for a minimum of one year. Given this, annual re-calibration is required for best performance.
  • Page 16: Specifications

    5. Specifications T/C Ranges: J, K, T, E, R, S, N, B, L, U including – 10 to 70 mV range RTD Ranges: Pt 385 (100, 200, 500, 1000 ohms) Pt 392, JIS, Ni 120, CU10, YSI 400 Ohms Ranges: 0 to 400.00 and 400.0 to 3200.0 T/C Type J: ±.3°C ±10 µV (±0.4°C total error) Accuracy:...
  • Page 17 Thermocouple Read and Source (All errors included) Range Accuracy B Thermocouple 2.1°C 1000 1.7°C 1000 1800 1.3°C L Thermocouple -200 0.45°C 900.0 0.4°C U Thermocouple -200 0.7°C 400.0 0.45°C N Thermocouple -200 1.1°C 1300.0 0.6°C mV Read/Source -10.00 75.00 0.015% ± 2 RTD Read and Source Range Accuracy...
  • Page 18: Warranty

    Any package not so marked will not be accepted and will be returned to the shipper. Martel will not be responsible for damage as a result of poor return packaging. Out of warranty repairs and recalibration will be subject to specific charges.