Martel BetaGauge PI User's Reference Manual

Martel BetaGauge PI User's Reference Manual

Reference class high accuracy digital pressure test gauge

Advertisement

BetaGauge PI & PI PRO
Reference Class
High Accuracy Digital
Pressure Test Gauge
User Reference Manual
This
document
includes information
for both the CSA and
CSA/ATEX certified
versions of the
BetaGauge PI.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BetaGauge PI and is the answer not in the manual?

Questions and answers

Summary of Contents for Martel BetaGauge PI

  • Page 1 BetaGauge PI & PI PRO Reference Class High Accuracy Digital Pressure Test Gauge User Reference Manual This document includes information for both the CSA and CSA/ATEX certified versions of the BetaGauge PI.
  • Page 3: Customer Service

    1. Introduction The BetaGauge PI Reference Class (PIR) combines the high accuracy of digital electronics with the convenience and ease of use of an analog test gauge. Accurate to ± 0.04% of reading, the BetaGauge PIR can be used as a calibration reference, or in any application where high accuracy pressure measurement is required.
  • Page 4: Hazard Location Information/Approvals

    Symbol Description Power OFF Power ON Earth ground Risk of Danger. Important information. Refer to manual. Battery Hazardous Voltage Conforms to ATEX requirements Certified by CSA as conforming to relevant Canadian and USA standards Conforms to relevant European Union directives. Wheeled bin, conforms to EC directive 2002/96/EC Hazard Location Information/Approvals Ex Hazardous Areas...
  • Page 5 (i.e. being dropped) it should be examined for any damage that may cause a safety concern. If in doubt please return the unit for evaluation to Martel Electronics. Please refer to the Customer Service Section for contact information.
  • Page 6: Operation

    3. Operation Power: The standard BetaGauge PIR is supplied with 3 AA batteries installed. If you purchased the optional 24Volt powered version, batteries are not installed. Connect a 24V power supply to the terminal block on the rear of the gauge, noting proper polarity. Do not install batteries when external power will be used.
  • Page 7: Supervisory Mode

    6. Set sample rate: This determines how often pressure is sampled and the display is updated. Choices are 0.5, 1, 3, and 10 samples/second. Note that 10/sec provides the fastest response time. 7. Set TARE. This allows you to set a constant offset value, which is then subtracted from the measured pressure.
  • Page 8: Normal Operation

    • Auto shutdown adjustment (enable/disable) • Damping settings (enable/disable) • Sample rate setting (enable/disable) • Tare setting (enable/disable) • Custom engineering units (set scale factor) 4. Use the CONFIG key to scroll through the above choices, and keys to lock and unlock features. Press CONFIG to continue scrolling through the parameters, pressing ENTER at any point saves your settings and returns the gauge to normal operation.
  • Page 9: Changing The Batteries

    Explosion hazard Batteries must only be changed in an area known to be non-hazardous. BetaGauge PI Grasp the face ring on the BetaGauge PIR, turn it approximately ¼ turn counterclockwise and remove. Note: For ATEX certified versions remove the two screws on the rear of the case to release the front plate.
  • Page 10: Specifications

    RS-232 Interface An RS-232 interface is standard on the BetaGauge PIR. Serial communication can be used for configuration, calibration, and to transfer measurement data from the gauge. For detailed specifications on the interface and software communication, see pages 14 and 15 in this manual. WARNING The RS-232 interface must not be used in hazardous areas.
  • Page 11 Media Compatibility All ranges except the compound range: Liquids and gases compatible with 316 stainless steel Compound range: clean, dry, non-corrosive gas Environmental Operating Temperature -10 °C to +55 °C (14°F to 131°F) Storage -20 °C to +70 °C (-4 °F to +158 °F) Humidity 10% to 95% RH Non-condensing Pollution Degree II Mechanical...
  • Page 12 Appendix 1: BetaGauge P P I I R R Calibration Procedure Overview Calibration adjustment of the BetaGauge PIR is performed electronically via internal software with the case closed. There are no mechanical adjustments; all calibration commands and adjustments are done via the keypad, using the display to guide the user through the calibration process.
  • Page 13 Procedure: Screens shown in this manual represent the displays shown with a 500 psi Gauge. The BetaGauge PIR will prompt the technician for the appropriate pressure at each calibration point. 500.00 CAL MODE Use the Pressure Standard to output 500.00 psi (100%). After the output has stabilized, press the ENTER key to continue.
  • Page 14 unit takes readings, the screen will show ———-. When the readings are complete the screen should look as shown below. 250.00 CAL MODE Use the Pressure Standard to output 250.00 psi (50%). After the output has stabilized, press the ENTER key to continue. As the unit takes readings, the screen will show ———-.
  • Page 15 0.00 CAL MODE Use the Pressure Standard to output 0.00 psi. After the output has stabilized, press the ENTER key to continue. As the unit takes readings, the screen will show ———-. When the readings are complete the screen should look as shown below. -12.00 CAL MODE Note: Only some ranges use vacuum calibration.
  • Page 17: List Of Commands

    9. BetaGauge PIR Serial Interface Instructions Initiating Communication The terminal communications can be setup using terminal communication software on a PC. The terminal settings need to be set as follows: • Bits per second: 9600 • Data bits: 8 • Parity: None •...
  • Page 18: Parameter Units

    HC_SI_OFF Turns SI mode off. HC_SI_ON Turns SI mode on. CAL_STORE Stores calibration data. HC_AUTO_OFF Turns auto shutdown off HC_AUTO_ON Turns auto shutdown on CUST_MULT? Sets the multiplier for the custom unit type STREAM_OFF Turns streaming data off STREAM_ON Turns streaming data on HC_TEMP? Same as TEMP? VAL?
  • Page 19: Error Codes

    Error Codes Error Description A non-numeric entry was received where it should be a numeric entry Too many significant digits entered Invalid units or parameter value received Entry is above the upper limit of the allowable range Entry is below the lower limit of the allowable range A required command parameter was missing An invalid pressure unit was received An unknown command was received...
  • Page 20: Warranty

    You can also call 1-800-821-0023 to have a form faxed. Martel will not be responsible for damage as a result of poor return packaging. Out of warranty repairs and recalibration will be subject to specific charges. Under no circumstances will Martel Electronics be liable for any device or circumstance beyond the value of the product.
  • Page 24 Tel: (603) 434-1433 800-821-0023 Fax: (603) 434-1653 Martel Electronics 3 Corporate Park Drive Derry, NH 03038 P/N 0200018 Rev. D 7/10...

This manual is also suitable for:

Betagauge pi pro

Table of Contents