Manual – LIV® IQ Oxygen

6. MR Compatibility.

The LIV IQ products include aluminum cylinders, regardless of cylinder size
and contents. Aluminum cylinders have no magnetic parts.
The integrated cylinder valve has a low residual magnetism, which means
that the pull forces of a 3-tesla magnet are barely perceptible.

7. Cleaning.

The user is not expected to clean the LIV IQ package routinely. The LIV IQ
package is a multi-use device and maintenance and cleanliness is
managed by Linde Healthcare.
The customer is recommended to keep the cylinder package clean, dry and
free from oil and grease as most malfunctions are caused by the intrusion
of particles, dust or other contamination.
If the outside surface of the package must be cleaned, use only a soft
cloth. The cloth may be dry or damp (with clean water).
If local disinfection procedures require that the LIV IQ must be cleaned, the
cleaning solutions used must be compatible with: brass, plastic materials of
components (including labels) and oxygen.

8. Service and Maintenance.

SN
The serial number of the product can be found on the back of the
valve, beneath the pressure outlet (if fitted) or blanking cap (if no
pressure outlet is fitted). The serial number consists of the year and month
of valve manufacture, two letters and five numeric digits.
In the event of a leakage or suspected defective LIV IQ package – do not try
to perform any repair or maintenance by yourself but return the product to
Linde Healthcare. Repair and maintenance activities must only be performed
by Linde Healthcare.
If a leak occurs, never try to tighten individual components or to stop the leak.
The LIV IQ must not be brought any nearer than 30 cm from the opening of
the magnet. Do not move the package during MRI scan, as the MR image
quality can be degraded.
Do not use cleaning solutions containing ammonia.
Do not immerse the product in water or any other liquids.
Do not expose the product to higher temperature than 149 °F, +65 °C (like
autoclave).
Do not pour any liquids into the user outlets.
When using the product in the hospital, the standards and practices of the
hospital's hygiene plan must be applied accordingly.
Do not use any cleaning solutions containing aldehyde,
ammonia or peroxide compounds.
O2 ISOV U12
0029
CGA540 WP2000PSI
201506 IH 00041
Non-certified repair and maintenance irrevocably leads to loss of the liability
from the manufacturer for the safety and functionality of the product.