Inspection Prior To Recharging - Wagner Meters Rapid RH 5.0 Manual

For the charging of easycare calcheck

Advertisement

Once the top chamber is fully charged (72 hours), separate
the top chamber from the bottom chamber. Store the bottom
chamber for future recharging processes. Take the top chamber
and insert it into a provided red rubber cap, then cover the
chamber with either an additional red rubber cap or a sensor,
if performing a calibration verification. This lessens the
potential for premature evaporation. If capped with a sensor,
for calibration verification, an additional amount of time
(overnight being optimal) is needed for the sensor to acclimate
and an accurate reading to be procured.
If you cap with a red rubber cap and then apply a sensor at
a later time, an additional amount of time (overnight being
optimal) is needed for the sensor to acclimate and an accurate
reading to be procured.
For more information on charge times and sensor handling,
please read the Best Practices document for the 5.0 reusable
sensors.

INSPECTION PRIOR TO RECHARGING

1. Over time, the EasyCare
CalCheck
chambers will dehydrate
®
and cease to provide a known 75%
RH range. They will need to be
recharged with water. Prior to this,
they will need to be inspected.
2. If capped with a red rubber cap,
remove and inspect chamber.
5