Wagner Meters Rapid RH 5.0 Manual

For the charging of easycare calcheck

Advertisement

Quick Links

Advertisement

loading

Summary of Contents for Wagner Meters Rapid RH 5.0

  • Page 2: Table Of Contents

    TABLE OF CONTENTS Introduction ..................... 2 Preparing ......................3 Initial Charging ....................3 Inspection Prior to Recharging ..............5 Recharging ......................6...
  • Page 3: Introduction

    INTRODUCTION The EasyCare CalCheck units contain a saltwater solution that ® emits a known 75% RH range. This allows a user to check their sensors as needed. The EasyCare CalCheck units need to be filled before first ® use and visually inspected before each use to ensure a proper amount of water remains in the top chamber.
  • Page 4: Preparing

    PREPARING Assemble the following tools and materials: • Tweezers (optional) • EasyCare CalCheck units ® • Pipettes • Container of water INITIAL CHARGING 1. Remove the EasyCare CalCheck units from the ® shipping tube. 2. Remove the red rubber cap, and separate the two chambers by pulling them apart.
  • Page 5 3. Fill the included pipette up to the limit line with water. Limit Line The pipette now contains one shot of water. 4. Add contents of pipette to the bottom chamber. Repeat once, adding a total of 3 shots of water to the bottom chamber. Caution: Do not overfill.
  • Page 6: Inspection Prior To Recharging

    Once the top chamber is fully charged (72 hours), separate the top chamber from the bottom chamber. Store the bottom chamber for future recharging processes. Take the top chamber and insert it into a provided red rubber cap, then cover the chamber with either an additional red rubber cap or a sensor, if performing a calibration verification.
  • Page 7: Recharging

    Note: The photos below show two chambers, which are dry. Notice the crystalline state of the salt with no free-flowing water. If the unit is dry, proceed with charging. Reminder: EasyCare CalCheck chambers will dry out sooner if ® left uncapped by a rubber cap or sensor. RECHARGING 1. If the above conditions in the “Inspection Prior to...
  • Page 8 For more information visit RapidRH.com or call our worldwide toll-free number (844) 808-8765...