Conrad Electronic SI-7004 Operating Instructions Manual page 15

Lcd panel-meter

Advertisement

Available languages

Available languages

The value determined by a reference measuring instrument is set
with the trimmer at the panel meter (probably covered) that the
panel meter shows the same value as the ,,reference unit". In the
respective measuring range (observe wiring and safety rules), 2 V
or 20 V or 200 V adjustment is made at 60% of the measuring
range (e.g. 20 V measuring range at approx. 12.5 VDC). The basic
measuring range of 200 mV has been adjusted ex factory.
4.4 Overflow Indication
If the input voltage exceeds the max. input limits or if a measu-
ring range is exceeded, ,,
turned on) for overflow is displayed instead of the value.
Technical Data and Accuracy
5.1 Technical Data
Display
Polarity Display
Overflow Indication . . . : ,,I" or ,,-I" is displayed
Max. Input Current DC/AC
without wiring . . . . . . . : 199.9 mVDC
Operating Principle
Max. Measuring Rate
Decimal Point Position
1"
or ,,-1" (when the negative sign is
3 l/2-digit Liquid Cystal Display (LCD),
maximum reading 1999
automatic ,,-" for negative input signal
Dual Slope
2
3 measurements per second
D 1 or D 2 or D 3 by short insulated
wire connections to ,,V+"
15