No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of A&D Company Ltd. The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice.
Page 2
5. Remove two screws and the RS-232C board from the indicator's rear panel. 6. Install the AD-4329A-03 option board on the rear panel from inside. Fasten the board to the rear panel from outside, using the two screws removed in step 5.
Page 3
For details on the setting procedure, refer to the AD-4329A instruction manual. Item Parameter Description f-47 0 RS-422 (Factory setting) RS-422 / RS-485 RS-485 f-47 1 Data format and commands are the same as those of RS-232C. For details, refer to the description of RS-232C in the AD-4329A instruction manual.
Connection example Note: Signal polarities (A, B) of some host computers may be reversed. SG is not available for some host computers. RS-422 RS-485 Terminating resistor Terminating resistor (May be built into the (May be built into the host computer.) host computer.) Host computer Host computer...