  • Page 1 D 351 IMPULSE HEAT-SEALER v.1.5...
  • Page 2: Table Of Contents

    Impulse sealer GIMA D 351 WARNING BEFORE USING THE MACHINE, READ CAREFULLY THIS MANUAL FOR A CORRECT USE IN ACCORDANCE WITH THE SAFE RULES. CONTENTS I SAFETY RULES page 1 II EXPECTED USE page 1 III CHARACTERISTICS page 2 IV MACHINE INSTALLATION...
  • Page 3: I Safety Rules

    GIMA s.p.a. is not responsible for direct and indirect damages derived from an incorrect use of the heat-sealer. EXPECTED USE The GIMA D 351 is an electrical impulse sealing machine for closing sterilization bags as per DIN 58953 P. 7 rules.
  • Page 4: Characteristics

    Impulse sealer GIMA D 351 III. CHARACTERISTICS • Sealable materials: paper/polypropilene/polyesther-tyvek/polyesther-polyethilene • sealing rules DIN 58953 P. 7 rules • Sealing length: mm. 350 • sealing width: mm. 8 • Electronic control of the sealing time from 0÷5 sec. • Incorporated cutting device.
  • Page 5: List Of The External Parts

    Impulse sealer GIMA D 351 LIST OF THE EXTERNAL PARTS pict. 1 1 : power supply cable 3 : identification data plate 2 : main switch 4 : rear cover fixing screws 5 : pivoted plain pict 2 cover pivoted plain knob...
  • Page 6: Instructions For Use

    Impulse sealer GIMA D 351 pict. 3 1 : adjustment opening bars ball grip 3 : lower bar adjustment screw locknuts 2 : lower fixing cover screws 4 : cutter blade INSTRUCTIONS FOR USE THE OPERATOR HAS TO BE TRAINED AND KNOW THE CONTENTS OF THIS...
  • Page 7 Impulse sealer GIMA D 351 THE SEALING TIME HAS TO BE SET ACCORDING TO THE SPECIFICATIONS OF THE BAG TO BE SEALED AND THE DESIRED OUTPUT. A comparative electronic control limits the heating of the resistance to avoid any overheating in case of fast sealing sequence.
  • Page 8 Impulse sealer GIMA D 351 SAFETY ELECTRONIC CONTROL In case of fault of the heating electronic control, the RED WARNING led (n° 4 pict. 4) will switch on and, in the same time, the power supply of the resistance will be interrupted .
  • Page 9: Maintenance

    Impulse sealer GIMA D 351 VII. MAINTENANCE IT CAN BE EFFECTUATED ONLY BY TRAINED TECHNICIANS ALWAYS UNPLUG THE POWER SUPPLY CABLE (n° 1 pict. 1) 1) machine opening procedure : To access top the inner components it is needed to remove the cover (n° pict. 2) of the machine.
  • Page 10 Impulse sealer GIMA D 351 do not loosen the two adjustement stands (n° 2 pict. 6) g) restore the opening of the pivoted plain to its work position : act through the ball grip (n°1 pict. 3) till to get the “h” gap = 8 mm. as shown in the picture n° 6 pict.
  • Page 11 Impulse sealer GIMA D 351 pict. 7 sealing bar sub-resistance band sealing bar fixing screw socket glass-PTFE fixing washer resistance fixing screw glass-PTFE protective cloth compensation spring resistance resistance supply cable 4) upper mobile bar protections replacement The protective cloth (n° 31 fig. 8 ) and the pressure rubber (n° 32 fig. 8 ) of the upper mobile bar (n°...
  • Page 12 Impulse sealer GIMA D 351 pict. 8 upper mobile bar pression rubber glass-PTFE protective cloth bar fixing screw 5) cutter blade replacement : The cutter group blade has a time-life . For the replacement proceed as follows : TAKE MUCH CARE NOT TO HURT YOURSELVES WITH THE KNIFE a) extract the cutter knob (n°...
  • Page 13 Impulse sealer GIMA D 351 6) line fuses replacement UNPLUG THE POWER SUPPLY CABLE (n°1 pict.1) The 4 A protection line fuses could be interrupted , if the green light of the main switch (n° 2 pict. 1) does not light and the machine does not work.
  • Page 14: List Of The Internal Parts

    Impulse sealer GIMA D 351 VIII. LIST OF THE INTERNAL PARTS pict. 11 main switch transformer electric supply socket electronic timer board line fuses box EMQ ant jamming filter lower sealing bar command panel board opening pivoted plain spring electromagnet & resistance start micro switch...
  • Page 15: Electrical Diagrams 230 Vac

    Impulse sealer GIMA D 351 ELECTRIC DIAGRAM line 230 Vac 50/60 Hz alarm led cycle starting micro switch (elettromagnet) command panel heating starting micro switch electromagnet strap resistance ant jamming filter electronic safety board flat cable 9 pins electronic timer board...
  • Page 16 Impulse sealer GIMA D 351 line 115 Vac 50/60 Hz alarm led cycle starting micro switch (elettromagnet) command panel heating starting micro switch electromagnet strap resistance ant jamming filter electronic safety board flat cable 9 pins electronic timer board fuses 6,3 AF...
  • Page 17: List Of The Spare Parts For The Normal Maintenance

    Impulse sealer GIMA D 351 LIST OF THE SPARE PARTS FOR THE NORMAL MAINTENANCE CODE POSITION DESCRIPTION n° pieces 20010 12 pict. 7 complete sealing bar 20001 24-30 pict. 7 protec. cloth for upper & lower bar 20002 25 pict. 7...
  • Page 18: Problems & Solutions

    Impulse sealer GIMA D 351 XII. PROBLEMS & SOLUTIONS PROBLEM CAUSE SOLUTION The machine does not function and the -The power cable is not correctly - Insert correctly the plug into the warning light of the main switch does insert into the wall socket or in the...
  • Page 19: Warranty Terms

    Impulse sealer GIMA D 351 XIII. WARRANTY TERMS The GANDUS’s heat-sealer are built to perform and they are guaranteed for 12 months after delivery. For the duration of the warranty, the manufacturer will replace parts or elements that, under his examination, should result defective for factory construction, error or faulty materials, but not the parts presenting normal wear, demonstrating incorrect use of the equipments or tampering.