Page 1
LOGGER S3541 Temperature and relative humidity logger with two 0-5V inputs Instruction Manual...
Page 2
Instruction manual for use of temperature and RH logger S3541 Logger is designed for measurement and record of ambient temperature, relative humidity and two signals 0-5Vdc. Measuring sensors of temperature and humidity are attached to logger. Measured values including calculated dew point temperature are displayed on the two-line LCD display and are stored in selectable time interval to internal nonvolatile memory.
Page 3
COM ADAPTER, for connection via USB port it is necessary to use USB ADAPTER. Connect adapter connector to proper computer port and plug the adapter to the guide slots on the side of logger. IE-LOG-S3541-04...
Page 4
After starting the logging with the 6 hour interval the first value is stored at that whole hour to perform the storing also at 00.00, i.e. at the beginning of the day. First storing is performed at 6.00,12.00, 18.00 or 00.00hour - at the hour from the above nearest to the logging IE-LOG-S3541-04...
Page 5
90%. Every 5 s display is automatically switched to displaying of other measured or calculated quantity. Logger now displays ambient temperature and dew point temperature (LCD line marked by symbol DP). IE-LOG-S3541-04...
Page 6
It is possible to select "memory alarm mode" when alarm is indicated permanently up to reset from the PC. Active alarm is indicated (if display is ON) by blinking of the value of proper quantity on the display and arrow symbol appears at the LCD upper IE-LOG-S3541-04...
Page 7
Plug magnet to guide slots from logger front side and wait approximately 1 s for decimal point appears right on LCD upper line. After appearance it is necessary immediately (till indication point is displayed) to remove magnet from guide slots and logger switches ON. IE-LOG-S3541-04...
Page 8
• put the lid back again and screw four screws. Be careful to have the rubber sealing properly in the slot and tighten the screws properly to ensure the water resistance of the instrument. IE-LOG-S3541-04...
Page 9
The IP rating is only IP20 and the IP rating of the logger (IP67) is not possible to apply. Connection to sensor signal. Power supply is located near the sensor: Connection to sensor signal. Power supply is located near the logger: IE-LOG-S3541-04...
Page 10
At this case sensor is possible to power e.g. from ac/dc adapter, connected via coaxial power connector to K0945 adapter. Supply voltage and the power of the applied ac/dc adapter depend on types of connected sensors. IE-LOG-S3541-04...