LOGGER S5141
Three channel 0-5 Vdc datalogger
with internal temperature sensor and display
Instruction Manual

Advertisement

loading
Need help?

Need help?

Do you have a question about the S5141 and is the answer not in the manual?

Questions and answers

Summary of Contents for Comet System S5141

  • Page 1 LOGGER S5141 Three channel 0-5 Vdc datalogger with internal temperature sensor and display Instruction Manual...
  • Page 2 Instruction manual for use of the S5141 logger The device is designed to measure and record signals from up to three external sensors with output 0 - 5 Vdc with the possibility to record simultaneously the ambient temperature measured by the built-in sensor.
  • Page 3 PC software to set logging parameters and other features. The PC program is common for all types of loggers Sxxxx – for type S5141 it is necessary to have installed version 4.0.17.0 or higher. Older program version can cause logger reconfiguration together with loss of proper function! Installed program version can be verified in menu Help / About program…...
  • Page 4 This should be taken into account when set the time of the automatic switching ON of the logger. The external sensor can be connected only when the logger is switched off, otherwise it may cause incorrect writing to MIN / MAX memory and to memory of the measured values. IE-LOG-S5141-01...
  • Page 5 – if it blinks, data memory is filled up to more than 90%. Every 5 s display is automatically switched to displaying of measured quantity. Logger now displays value of 3.channel (LCD upper line) and ambient temperature (LCD lower line). IE-LOG-S5141-01...
  • Page 6 – logger would be switched OFF (memory of the MIN/MAX values would be erased)! Display reading initialized by magnet goes out automatically after 30 s. Remove magnet from slots anytime during actual reading is ON or later. Temporary displaying of actual reading by magnet IE-LOG-S5141-01...
  • Page 7 If reading BAT is displayed permanently, battery voltage is low and logger is not possible to switch ON. If logger was switched ON before it, data logging is stopped and logger is switched OFF. Communication with computer can temporarily work. Replace battery as soon as possible! IE-LOG-S5141-01...
  • Page 8 LCD upper line. After decimal point appearance it is necessary immediately (till indication point is displayed) to remove magnet from guide slots. Reading CLR MIN MAX appears for several seconds and MIN/MAX values will be cleared. IE-LOG-S5141-01...
  • Page 9 EN 61000-4-3 (intensity of electric field 3 V/m, Class A) EN 61000-4-4 (levels 1/0.5 kV, Class A) EN 61000-4-6 (intensity of electric field 3 V/m, Class A) Technical support and service Technical support and service is provided by distributor. Contact is included in warranty certificate. IE-LOG-S5141-01...