(30) Trigger MODE switch
(31) Trigger COUPLING switch
(32) Trigger SOURCE switch
To select the sweep triggering mode.
AUTO position selects free-running sweep where
a baseline is displayed in the absense of a
signal. This condition automatically reverts
to triggered sweep when a trigger signal of 25
Hz
or higher is received and other trigger
controls are properly set.
NORM position produces sweep only when a trigger
signal is received and other controls are pro­
perly set. No trace is visible if any trigger
requirement is missing. This mode must be used
when the signal frequency is 25
Hz
or lower.
TV-V position is used for observing a composite
video signal at the frame rate.
TV-H position is used for observing a composite
video signal at the line rate.
To select the frequency characteristics of the
trigger-circuit coupling.
AC position inserts a large capacitor in the
trigger-coupling chain to remove and DC com­
ponent from the trigger signal.
HF
REJ
position inserts a filter in the trigger­
coupling chain that removes signal components
higher than 4 kHz. Use this position to elimin­
ate high-frequency noise.
LF
REJ
position inserts a filter that removes
signal components lower in frequency than 4 kHz.
Use this position to eliminate low-frequency
noise or undesired trigger signals.
DC position passes the trigger signal without
modification. Use this setting when triggering
from a very low-frequency signal or a DC level.
To conveniently select the trigger source.
CHI position selects the channel I signal as
the trigger source.
CH2 position selects the channel 2 signal as
the trigger source.
LINE position selects a trigger derived from the
AC power line. This permits the scope to sta­
bilize display line-related components of a signal
even if they are very small compared to other
components of the signaL
EXT position selects the signal applied to the
EXT TRIG IN connector.