Title: CWPC_GLU_0140 Accu-Chek Inform II Training Guide
Note: Lot numbers for QC or Test Strips cannot be manually entered.
5. "Flashing Arrow" symbol appears. Insert test strip into test strip port at top
of meter.
6. "Flashing Drop" symbol appears.
 Meter is ready for QC sample
 Mix the control vials by gently inverting 5 times
 Remove cap and wipe vial with tissue
 Discard first drop and apply 2nd drop to FRONT EDGE of strip.
Note: Test strip must ALWAYS be in meter BEFORE application of QC solution
Position the meter on a flat surface or downwards – DO NOT allow liquid in
test strip port.
Blue dye in solution can stain clothing
Result will appear within 5 seconds.
7. Touch "RANGE" "OUT OF RANGE" symbol to view the QC range.
If Result in range, touch "Check Mark" to return to Main Menu screen.
If result is out of reference range for level, 'Out of Range' symbol will flash.
A comment is required when QC is "Out of Range" or if there is an
"ERROR CODE".
 Touch "Cloud" symbol in lower left corner of screen.
 Pre-programmed comments appear to select.
can be selected)
 Touch "Checkmark" in lower right to confirm selection.
 Touch "Check Mark" again if satisfied with comment to exit from
screen.
Note: Once exited from QC result/comment screen, new comment cannot be
added later
Pre-Programmed Comments for Quality Control Testing available to use in Meters
@ PHSA:
 QC Procedure Error
 Wrong QC Vial
 To Repeat QC Test
 To Remix QC vial
 Repeated QC Test
 Open new Strip Vial
Patient Test - Puncture Site & Sample Collection
 Puncture site is clean. Clean site with soap and water if required.
 Disinfectant is 2% Chlorhexidine & 70% isopropyl alcohol
Medical Approval: Dr Benjamin Jung
Version: 1.2
Folder Name: CW\Point of Care\Whole Blood Glucose
This is a controlled document for CW use only. Any printed copies are uncontrolled unless specified. Please refer to Lab QMS
PHSA Laboratories CW Site - Point of Care